8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2021-02878
- Event Type
- Malfunction
- Date Received
- January 7, 2021
- Report Date
- January 22, 2020
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403810015
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. THE DEVICE WAS REPAIRED, RETESTED, AND RETURNED TO THE CUSTOMER. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
ALARM - ERROR CODES / MESSAGES LOGIC BOARD- CHIP PIN DAMAGE AND RECALL AFFECTED ALARM - ERROR CODES / MESSAGES- 01/22/2020 05:29:37 RFC_REPAIRS (RFC_REPAIRS) (B)(4). DOWNLOADED ERROR LOG. CONTAINS NUMEROUS CODES. TECH SERVICE RECOMMENDED REPLACING MOTOR CONTROL AND POWER SUPPLY BOAR DS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30277 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 | 10885403810015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |