FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAGSLA

MDR report key: 11127861 · Received January 6, 2021

Report

Report Number
1920898-2020-01800
Event Type
Malfunction
Date Received
January 6, 2021
Date of Event
December 9, 2020
Report Date
March 12, 2021
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY : NO SAMPLES WERE RETURNED THEREFORE THE INVESTIGATION WAS PERFORMED BASED ON THE PHOTOS PROVIDED. ONE PHOTO OF 1 LOOSE 0.3ML BD INSULIN SYRINGE WAS PROVIDED. THE CUSTOMER REPORTED WHEN HE REMOVES THE SAFETY SHIELD (ORANGE COVER), THE NEEDLE LOOSES FROM THE SYRINGE BODY (TOGETHER WITH THE PLASTIC PART), GETTING STUCK IN THE ORANGE SHIELD. THE PHOTO WAS EXAMINED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS SEPARATED FROM THE BARREL. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231312. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA1630423 HAS BEEN OPENED TO ADDRESS THIS ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT (B)(4) SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAGSLA SEPARATED FROM THE HUB DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IN THE LAST BATCH PURCHASED, WHEN HE REMOVES THE SAFETY SHIELD (ORANGE COVER), THE NEEDLE LOOSES FROM THE SYRINGE BODY (TOGETHER WITH THE PLASTIC PART), GETTING STUCK IN THE ORANGE SHIELD. IT OCCURRED IN (B)(4) SYRINGES FROM THE SAME BATCH. HAS OCCURRED IN 2017 TOO."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. INITIAL REPORTER PHONE: (B)(6). (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT 3 SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAGSLA SEPARATED FROM THE HUB DURING USE. THE FOLLOWING WAS REPORTED BY THE INITIAL REPORTER: "IN THE LAST BATCH PURCHASED, WHEN HE REMOVES THE SAFETY SHIELD (ORANGE COVER), THE NEEDLE LOOSES FROM THE SYRINGE BODY (TOGETHER WITH THE PLASTIC PART), GETTING STUCK IN THE ORANGE SHIELD. IT OCCURRED IN 3 SYRINGES FROM THE SAME BATCH. HAS OCCURRED IN 2017 TOO."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19892 SYRINGE 0.3ML 31GA 6MM HALFUNIT 10BAGSLA PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 9231312

Patients

Seq Age Sex Outcome Treatment
1 50 YR