FDA Adverse Event Malfunction Summary report: N

ARCHITECT C16000

MDR report key: 11127805 · Received January 6, 2021

Report

Report Number
3016438761-2021-00007
Event Type
Malfunction
Date Received
January 6, 2021
Date of Event
December 14, 2020
Report Date
February 26, 2021
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740005924
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

SECTION A: SAMPLE ID ADDED, (B)(6). THE FIELD SERVICE REPRESENTATIVE (FSR) PERFORMED TROUBLESHOOTING AND REPLACED THE BELLOWS, BELLOWS ONLY (ROHS) PART NUMBER 2-89054-02 DUE TO LEAKING. A REVIEW OF SERVICE HISTORY FOR THE ARCHITECT C16000 SERIAL NUMBER (B)(6) REVEALED NO ADDITIONAL DISCREPANT RESULTS REPORTED, NOR DID IT IDENTIFY ANY CONTRIBUTING FACTORS TO THE CURRENT COMPLAINT. A REVIEW OF TRACKING AND TRENDING FOR THE BELLOWS, BELLOWS ONLY (ROHS) PART NUMBER 2-89054-02 DID NOT IDENTIFY ANY TRENDS. REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT C16000 DID NOT IDENTIFY ANY TRENDS. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE INSTRUMENT PART OR THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE INVESTIGATION NO SYSTEMIC ISSUE OR DEFICIENCY OF THE BELLOWS, BELLOWS ONLY (ROHS) PART NUMBER 2-89054-02 OR THE ARCHITECT C16000 SERIAL NUMBER (B)(6) WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. NO PATIENT INFORMATION IS AVAILABLE.

Description of Event or Problem · 1

THE CUSTOMER GENERATED FALSELY ELEVATED ENZYMATIC CREATININE RESULTS WHILE USING THE ALINITY C ENZYMATIC CREATININE (LIST NUMBER LN 8L24-31). THE FOLLOWING DATA WAS PROVIDED: INITIAL RESULT REPORTED OUT OF THE LAB, 606 UMOL/L, REPEAT 71.7 UMOL/L AND REPEAT AT ANOTHER FACILITY 70 UMOL/L NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23941 ARCHITECT C16000 ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 3L77-01 00380740005924

Patients

Seq Age Sex Outcome Treatment
1 MLTGNT ENZY CREAT, 08L24-31, 00028Y600| MLTGNT ENZY CREAT, 08L24-31, 00028Y600