ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) PISTOL GRIP 36CM-15MM ACTIVE BL
Report
- Report Number
- 3005075853-2008-01090
- Event Type
- Malfunction
- Date Received
- August 11, 2008
- Date of Event
- May 25, 2008
- Report Date
- June 26, 2008
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K002981
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
DATE SENT: 08/11/2008. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WILL STOP ACTIVATING, AND EITHER EMIT A SOLID TONE OR DISPLAY AN INSTRUMENT ERROR CODE 5 ON THE GENERATOR DISPLAY WHEN THE BLADE BECOMES DAMAGED. THE DEVICE WAS TESTED WITH A GEN04 AND AN ERROR CODE 5 WAS NOTED. WHEN A BLADE HAS BEEN COMPROMISED, SCRATCHED OR CRACKED, THE TITANIUM METAL IS FATIGUED WHEN CONTINUALLY ENERGIZED AND THIS RESULTS IN THE BLADE FURTHER CRACKING AND POSSIBLE BREAKING OFF. DURING THE ANALYSIS PROCESS, THE BLADE WAS TESTED FOR SCRATCHES AND CRACKS AND THE BLADE FURTHER CRACKED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. THEREFORE, THE INSTRUCTIONAL INSERT STATES: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT, AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED AT THIS PROCESS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A CARDIOPLASTY PROCEDURE, THE DEVICE STOPPED WORKING. UNKNOWN HOW CASE WAS COMPLETED. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL CURVED SHEARS (LCS) PISTOL GRIP 36CM-15MM ACTIVE BL | GEI | ETHICON ENDO-SURGERY, LLC. | NA | D4JT94 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR| HANDPIECE |