FDA Adverse Event Malfunction Summary report: N

SIMPLEX P TWIN PACK

MDR report key: 1112509 · Received August 7, 2008

Report

Report Number
9610726-2008-00049
Event Type
Malfunction
Date Received
August 7, 2008
Date of Event
July 17, 2008
Report Date
July 24, 2008
Manufacturer
STRYKER ORTHOPAEDICS LIMERICK
Product Code
LOD
PMA / PMN Number
N17004
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION OF THE DEVICE CANNOT BE PERFORMED AS THE DEVICE WAS NOT RETURNED TO THE MFR. IF ADDITIONAL INFO BECOMES AVAILABLE, IT WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

DURING SURGERY, THE CEMENT SET TOO QUICKLY. IT SET IN 6 TO 7 MINUTES ALTHOUGH USUALLY 11 TO 12 MINUTES. THE SURGEON IS VERY EXPERIENCED WITH THE SIMPLEX CEMENT AND HE ADDRESSED THAT HE KEPT AND HANDLED THE CEMENT AS USUAL, HOWEVER, IT SET TOO QUICKLY. OPERATING ROOM TEMPERATURE WAS AT 24C. DISTRIBUTOR'S WAREHOUSE TEMPERATURE WAS 25C DURING THE DAY, HOWEVER, THE AIR CONDITIONER WAS OFF DURING NIGHTS. NIGHT IS AROUND 30C OR HIGHER, HOWEVER, IT GETS VERY WARM DURING SUMMER TIME. THE CEMENT WAS BROUGHT TO THE HOSP THE DAY BEFORE THE SURGERY AND KEPT IN A REFRIGERATOR UNTIL 10 MINUTES BEFORE THE SURGERY. THE MIXING WAS DONE MANUALLY. NO ANTIBIOTIC WAS MIXED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SIMPLEX P TWIN PACK IMPLANT LOD STRYKER ORTHOPAEDICS LIMERICK NA JDP018

Patients

Seq Age Sex Outcome Treatment
1 Other