SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS
Report
- Report Number
- 1920898-2021-00003
- Event Type
- Malfunction
- Date Received
- January 6, 2021
- Date of Event
- December 8, 2020
- Report Date
- January 26, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382908468039
- PMA / PMN Number
- K024112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 12/28/2020 . H.6. INVESTIGATION: CUSTOMER RETURNED (3) 1/2CC, 8MM, 31G SYRINGES IN AN OPEN POLY BAG FROM LOT # 0174632. CUSTOMER STATES THAT THE HUB ASSEMBLY REMAINED IN THE NEEDLE SHIELDS AND THAT THE SYRINGES HAD THE SHIELDS LOOSE IN THE BAG. ALL RETURNED SYRINGES WERE EXAMINED AND ALL EXHIBITED A BROKEN BARREL TIP, WHICH WILL BE INVESTIGATED UNDER INVESTIGATION CHILD 2160821. NO HUB ASSEMBLY SEPARATION WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 0174632. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE TWO (2) NOTIFICATIONS [200898951, 200898499] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THERE WAS ONE (1) NOTIFICATIONS [200897792] NOTED FOR HIGH SHIELD PULLS. ROOT CAUSE: L2L #103316 DISPATCH CORRECTED THE ISSUE BY OILING THE SLIDES GOING INTO THE FIRST INDEX DIAL.
IT WAS REPORTED THAT SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS NEEDLE HUB SEPARATED ON 3 OCCASION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 328468 BATCH NO.: 0174632. IT WAS REPORTED THAT THE HUB ASSEMBLY REMAINED IN THE NEEDLE SHIELDS AND THREE SYRINGES HAD THE SHIELDS LOOSE IN THE BAG. PER SNOW: CONSUMER REPORTED FOUND 3 SYRINGE IN THIS ONE BAG NEEDLE SHIELDS WITH NEEDLE HUB ASSEMBLY INSIDE SHIELD, AND JUST LAYING IN THE BAG. THREE SYRINGE BARRELS WITHOUT THE NEEDLE SHIELDS ALSO IN THE BAG.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS NEEDLE HUB SEPARATED ON 3 OCCASION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 328468, BATCH NO.: 0174632. IT WAS REPORTED THAT THE HUB ASSEMBLY REMAINED IN THE NEEDLE SHIELDS AND THREE SYRINGES HAD THE SHIELDS LOOSE IN THE BAG. PER (B)(6): CONSUMER REPORTED FOUND 3 SYRINGE IN THIS ONE BAG NEEDLE SHIELDS WITH NEEDLE HUB ASSEMBLY INSIDE SHIELD, AND JUST LAYING IN THE BAG. THREE SYRINGE BARRELS WITHOUT THE NEEDLE SHIELDS ALSO IN THE BAG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 19545 | SYRINGE 0.5ML 31GA 8MM UFII 10BAG 500CS | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 328468 | 0174632 | 00382908468039 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |