BD MICRO-FINE IV INSULIN SYRINGE
Report
- Report Number
- 1920898-2020-01798
- Event Type
- Malfunction
- Date Received
- January 5, 2021
- Date of Event
- December 9, 2020
- Report Date
- March 5, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 30382903294245
- PMA / PMN Number
- K170386
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D10: DEVICE AVAILABLE FOR EVAL YES. D10: RETURNED TO MANUFACTURER ON: 2021-02-22. INVESTIGATION SUMMARY: CUSTOMER RETURNED (3) 1CC, 12.7MM, 28G SYRINGES IN BLISTER PACKS FROM LOT # 9119537, 9322637, 0160417, 0099954, 9294983, 9231122, 9196270, 9189788, 9168926, 9140521, 8351757, AND 8302667. CUSTOMER STATES THAT THE PACKAGING WAS NOT SEALED ON ONE SIDE. ALL RETURNED SYRINGES WERE EXAMINED AND ALL BLISTER PACKS WERE OBSERVED TO BE OPEN. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9119537, 9322637, 0160417, 0099954, 9294983, 9287332, 9231122, 9196270, 9189788, 9168926, 9140521, 8351757, AND 8302667. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT.
IT WAS REPORTED THAT SEVERAL BD MICRO-FINE¿ IV INSULIN SYRINGES HAD ISSUES WITH PARTIAL UNSEALED PACKAGING UNITS. THERE WERE 3 FROM LOT 9119537, 14 FROM 9140521, 18 FROM 9168926, 30 FROM 9189788, 25 FROM 9196270, 5 FROM 9231122, 1 FROM 9287332, 38 FROM 9294983, 15 FROM 0099954, 256 FROM 0160417, 23 FROM 8302667, 36 FROM 8351757, AND 3 FROM 9322637. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ITEM 329424, A TOTAL OF 467 UNITS ACROSS 13 LOT NUMBERS WERE FOUND TO BE PARTLY UNSEALED"
THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9119537. MEDICAL DEVICE EXPIRATION DATE: 2024-07-31. DEVICE MANUFACTURE DATE: 2019-04-29. MEDICAL DEVICE LOT #: 9140521. MEDICAL DEVICE EXPIRATION DATE: 2024-08-31. DEVICE MANUFACTURE DATE: 2019-05-20. MEDICAL DEVICE LOT #: 9168926. MEDICAL DEVICE EXPIRATION DATE: 2024-09-30. DEVICE MANUFACTURE DATE: 2019-06-17. MEDICAL DEVICE LOT #: 9189788. MEDICAL DEVICE EXPIRATION DATE: 2024-09-30. DEVICE MANUFACTURE DATE: 2019-07-08. MEDICAL DEVICE LOT #: 9196270. MEDICAL DEVICE EXPIRATION DATE: 2024-10-31. DEVICE MANUFACTURE DATE: 2019-07-15. MEDICAL DEVICE LOT #: 9231122. MEDICAL DEVICE EXPIRATION DATE: 2024-11-30. DEVICE MANUFACTURE DATE: 2019-08-19. MEDICAL DEVICE LOT #: 9287332. MEDICAL DEVICE EXPIRATION DATE: 2024-12-31. DEVICE MANUFACTURE DATE: 2019-10-14. MEDICAL DEVICE LOT #: 9294983. MEDICAL DEVICE EXPIRATION DATE: 2024-12-31. DEVICE MANUFACTURE DATE: 2019-10-21. MEDICAL DEVICE LOT #: 0099954. MEDICAL DEVICE EXPIRATION DATE: 2025-06-30. DEVICE MANUFACTURE DATE: 2020-04-08. MEDICAL DEVICE LOT #: 0160417. MEDICAL DEVICE EXPIRATION DATE: 2025-08-31. DEVICE MANUFACTURE DATE: 2020-06-08. MEDICAL DEVICE LOT #: 8302667. MEDICAL DEVICE EXPIRATION DATE: 2024-01-31. DEVICE MANUFACTURE DATE: 2018-10-29. MEDICAL DEVICE LOT #: 8351757. MEDICAL DEVICE EXPIRATION DATE: 2024-03-31. DEVICE MANUFACTURE DATE: 2018-12-17. MEDICAL DEVICE LOT #: 9322637. MEDICAL DEVICE EXPIRATION DATE: 2025-02-28. DEVICE MANUFACTURE DATE: 2019-11-18. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SEVERAL BD MICRO-FINE¿ IV INSULIN SYRINGES HAD ISSUES WITH PARTIAL UNSEALED PACKAGING UNITS. THERE WERE 3 FROM LOT 9119537, 14 FROM 9140521, 18 FROM 9168926, 30 FROM 9189788, 25 FROM 9196270, 5 FROM 9231122, 1 FROM 9287332, 38 FROM 9294983, 15 FROM 0099954, 256 FROM 0160417, 23 FROM 8302667, 36 FROM 8351757, AND 3 FROM 9322637. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "ITEM 329424, A TOTAL OF 467 UNITS ACROSS 13 LOT NUMBERS WERE FOUND TO BE PARTLY UNSEALED"
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17854 | BD MICRO-FINE IV INSULIN SYRINGE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 329424 | SEE SECTION H.10. | 30382903294245 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |