FDA Adverse Event Malfunction Summary report: N

VERSACARE

MDR report key: 1112085 · Received August 15, 2008

Report

Report Number
1824206-2008-00475
Event Type
Malfunction
Date Received
August 15, 2008
Date of Event
March 23, 2007
Report Date
March 23, 2007
Manufacturer
HILL-ROM RITTER
Product Code
FNL
Removal / Correction Number
Z-0240-2008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON COMPLAINT REVIEW, IT WAS DETERMINED THAT IS A REPORTABLE EVENT FOR SIDERAIL NOT LATCHING. REPLACEMENT LATCH KIT INSTALLED WHICH RESOLVED THE PROBLEM.

Description of Event or Problem · 1

BED SIDERAILS WOULD NOT LATCH. THERE WERE NO INJURIES IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VERSACARE AC POWERED HOSPITAL BED FNL HILL-ROM RITTER P3200 NA

Patients

Seq Age Sex Outcome Treatment
1