OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2008-00166
- Event Type
- Malfunction
- Date Received
- August 5, 2008
- Date of Event
- July 9, 2008
- Report Date
- July 9, 2008
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
THE PRODUCT WAS NOT RETURNED SO NO EVALUATION IS POSSIBLE. THE CUSTOMER NOTICED A "CLICKING" SOUND DURING THE FILL PROCESS WHICH INDICATES A PROBABLE PROBLEM WITH THE RETAINER THAT ALLOWED A COMPONENT TO MOVE RESULTING IN DAMAGE. THIS DAMAGE PREVENTED THE PLUNGER FROM ADVANCING. THE USER WOULD HAVE BEEN AWARE OF A PROBLEM DURING THE FILL PROCESS. THERE IS RESISTANCE FELT IN FILLING THE POD WITH INSULIN AND A DISTINCT "CRACKLING" NOISE RESULTING FROM STRIPPING THE THREADS OF THE COMPONENT WHEN OVER-PRESSURIZED. THE OMNIPOD USER GUIDE STATES: "WARNING: NEVER USE A POD IF, DURING FILL, YOU DETECT ANY CRACKLING NOISE OR RESISTANCE WHILE DEPRESSING THE PLUNGER OF THE FILL SYRINGE. USING A POD WITH THESE CONDITIONS COULD RESULT IN UNDER-DELIVERY OF INSULIN." THE USER IS ALSO INSTRUCTED IN THE USER GUIDE TO MONITOR BLOOD GLUCOSE LEVELS FREQUENTLY. BY FOLLOWING THESE RECOMMENDATIONS, THE USER BECAME AWARE OF THEIR HIGH BLOOD GLUCOSE AND STARTED A NEW POD.
CUSTOMER CALLED IN TO REPORT THAT HE HAD CLICKING NOISES IN HIS POD WHEN HE FILLED IT WITH INSULIN. CUSTOMER WORE POD PUT THEN HE STARTED TO HAVE HIGH BLOOD GLUCOSE LEVELS. CUSTOMER REMOVED POD AND WAS ABLE TO ACTIVATE A NEW POD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L11683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |