FDA Adverse Event Malfunction Summary report: N

PROGRAMMING WAND

MDR report key: 1111875 · Received August 12, 2008

Report

Report Number
1644487-2008-01875
Event Type
Malfunction
Date Received
August 12, 2008
Date of Event
July 23, 2008
Report Date
July 23, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

REPORTER INDICATED A VNS WAND FAILED TO PROGRAM PTS. THE VNS WAND WAS RETURNED FOR PRODUCT ANALYSIS. ANALYSIS REVEALED THAT THE WAND SERIAL DATA CABLE, WHICH PRODUCED COMMUNICATION ERRORS, HAD AN OPEN CONDUCTOR. A KNOWN GOOD BENCH SERIAL DATA CABLE WAS SUBMITTED AND ALL COMMUNICATIONS ERRORS CLEARED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING WAND PROGRAMMING WAND LYJ CYBERONICS, INC. 201

Patients

Seq Age Sex Outcome Treatment
1