FDA Adverse Event
Malfunction
Summary report: N
KIMBERLY CLARK-BALLARD MEDICAL R-2 PAD ADAPTOR
MDR report key: 1111809
·
Received August 13, 2008
Report
- Report Number
- MW5007977
- Event Type
- Malfunction
- Date Received
- August 13, 2008
- Date of Event
- July 17, 2008
- Report Date
- August 12, 2008
- Manufacturer
- CON MED
- Product Code
- MKJ
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
PACER CABLE WITH KIMBERLY CLARK - BALLARD MEDICAL R-2 ADAPTOR IN LINE DID NOT FUNCTION PROPERLY DURING CARDIAC ARREST. THERE WAS A DELAY OF LESS THAN TWO MINUTES WHILE A SECONDARY CABLE WAS SECURED. EVALUATION BY CLINICAL ENGINEERING DETERMINED THAT THE PAD ADAPTER - THAT WAS SECURED TO THE PACER CABLE TO ALLOW USE OF A SPECIFIC PAD - WAS FAULTY CAUSING THE CABLE TO BE NON-OPERATIONAL. X-RAY OF ADAPTOR REVEALED FRACTURED WIRES IN ADAPTOR. DELAY DID NOT EFFECT PATIENT OUTCOME. TWO ADDITIONAL KIMBERLY CLARK - BALLARD MEDICAL R-2 ADAPTORS WERE FOUND TO BE NON-FUNCTIONAL DURING ROUTINE PREVENTIVE MAINTENANCE CHECKS. THEY WERE NOT INVOLVED WITH PATIENT CARE. UNABLE TO DETERMINE LOT NUMBERS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KIMBERLY CLARK-BALLARD MEDICAL R-2 PAD ADAPTOR | DEFIBRILLATOR PAD ADAPTOR | MKJ | CON MED | 32-170555 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |