FDA Adverse Event Malfunction Summary report: N

KIMBERLY CLARK-BALLARD MEDICAL R-2 PAD ADAPTOR

MDR report key: 1111809 · Received August 13, 2008

Report

Report Number
MW5007977
Event Type
Malfunction
Date Received
August 13, 2008
Date of Event
July 17, 2008
Report Date
August 12, 2008
Manufacturer
CON MED
Product Code
MKJ
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

PACER CABLE WITH KIMBERLY CLARK - BALLARD MEDICAL R-2 ADAPTOR IN LINE DID NOT FUNCTION PROPERLY DURING CARDIAC ARREST. THERE WAS A DELAY OF LESS THAN TWO MINUTES WHILE A SECONDARY CABLE WAS SECURED. EVALUATION BY CLINICAL ENGINEERING DETERMINED THAT THE PAD ADAPTER - THAT WAS SECURED TO THE PACER CABLE TO ALLOW USE OF A SPECIFIC PAD - WAS FAULTY CAUSING THE CABLE TO BE NON-OPERATIONAL. X-RAY OF ADAPTOR REVEALED FRACTURED WIRES IN ADAPTOR. DELAY DID NOT EFFECT PATIENT OUTCOME. TWO ADDITIONAL KIMBERLY CLARK - BALLARD MEDICAL R-2 ADAPTORS WERE FOUND TO BE NON-FUNCTIONAL DURING ROUTINE PREVENTIVE MAINTENANCE CHECKS. THEY WERE NOT INVOLVED WITH PATIENT CARE. UNABLE TO DETERMINE LOT NUMBERS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KIMBERLY CLARK-BALLARD MEDICAL R-2 PAD ADAPTOR DEFIBRILLATOR PAD ADAPTOR MKJ CON MED 32-170555 UNK

Patients

Seq Age Sex Outcome Treatment
1 50 YR