FDA Adverse Event Injury Summary report: N

THE SIMPLE LOCKING INTRA-MEDULLARY (SLIM) SYSTEM

MDR report key: 11117832 · Received January 5, 2021

Report

Report Number
3000327445-2021-00001
Event Type
Injury
Date Received
January 5, 2021
Report Date
January 4, 2021
Manufacturer
PEGA MEDICAL INC.
Product Code
HSB
UDI-DI
07540194002662
PMA / PMN Number
K192710
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE BONE CANCER CONTRIBUTED TO THE NON-UNION OF THE BONE. ALSO, IT IS IMPORTANT TO NOTE THAT NONUNION IS PREVALENT IN THE FOREARM DUE TO THE DIFFICULTY IN MAINTAINING FULL IMMOBILITY. ARM MOVEMENT DURING NORMAL DAILY ACTIVITY EXPOSES THE HUMERUS AND FOREARM TO LARGE BENDING FORCES AS THEY ACT AS LEVERS. THEREFORE, IT IS OFTEN DIFFICULT TO ACHIEVE BONE UNION, ESPECIALLY WHEN THE BONE DEFECT IS LARGE WITH CANCEROUS BONE, AS IN THIS CASE. THE NONUNION CAUSED FATIGUE FAILURE OF THE EXPOSED SLIM ROD DUE TO CYCLIC CANTILEVER FORCES AT THE FRACTURE SITE. THE FACT THAT THERE IS NO VISIBLE PLASTIC DEFORMATION ON THE IMPLANT CONFIRMS THE FATIGUE FAILURE. FURTHERMORE, REVIEW OF THE MANUFACTURING FILES SHOWED THAT ALL DIMENSIONAL INSPECTIONS AND MATERIAL MECHANICAL PROPERTIES WERE WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

A SLIM ROD INSERTED INTO THE RIGHT ULNA OF A 6-YEAR-OLD PATIENT BROKE AT THE NONUNION SITE. THE NAIL WAS IMPLANTED ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11819 THE SIMPLE LOCKING INTRA-MEDULLARY (SLIM) SYSTEM SLIM ROD Ø2.0MM X 140MM HSB PEGA MEDICAL INC. SLM-20-140 160405-007 07540194002662

Patients

Seq Age Sex Outcome Treatment
1 6 YR Required Intervention