TALENT ABDOMINAL BIFURCATED STENT GRAFT SYSTEM

MDR report key: 1111731 · Received August 14, 2008

Report

Report Number
2953200-2008-00676
Event Type
Injury
Date Received
August 14, 2008
Date of Event
July 17, 2008
Report Date
July 17, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SECONDARY INTERVENTION REQUIRED. EVAL RESULTS: (OCCLUSION), (SEVERE TORTUOSITY OF THE ILIAC LIMB). CONCLUSION: (SEVERE TORTUOSITY OF THE ILIAC LIMB).

Description of Event or Problem · 1

A TALENT STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS SEVERE TORTUOSITY, MILD CALCIFICATION AND MILD ANGULATION OF THE SHORT AORTIC NECK. IT WAS REPORTED THAT THE CONTRA LATERAL LIMB WAS BOWING/CURVING AROUND. IT WAS REPORTED TWO DAYS AFTER DEPLOYMENT OF THE IPSILATERAL LIMB STENT GRAFT WAS DEPRESSED DUE TO THE CONTRA LIMB PUSHING OVER ON THE CONTRA LIMB WHICH WAS UNSUPPORTED. THE PHYSICIAN ELECTED TO PLACE TWO ILIAC LIMBS, ONE LIMB ON EACH SIDE FOR REINFORCEMENT OF THE STENT GRAFT. THE PHYSICIAN BELIEVES THE SEVERE TORTUOSITY OF THE ILIAC LIMB RESULTED IN THE ILIAC LIMB BOWING CAUSING VESSEL OCCLUSION. NO ADD'L CLINICAL SEQUELAE WERE REPORTED AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL BIFURCATED STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA V00114523

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention