FDA Adverse Event
Malfunction
Summary report: N
SYNERGY
MDR report key: 1111636
·
Received August 13, 2008
Report
- Report Number
- 3004209178-2008-04916
- Event Type
- Malfunction
- Date Received
- August 13, 2008
- Date of Event
- July 5, 2008
- Report Date
- July 15, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
IMPEDANCES ON ALL OR SOME OF THE BIPOLAR PAIRS WERE GREATER THAN 4000 OHMS. ALL OTHER IMPEDANCES WERE 932. THE PT HAD RELIEF FROM THERAPY. THE DEVICE AMPLITUDE HAD BEEN SET AT 10.5 VOLTS AT TIMES. THE BATTERY MEASURED 2.56 VOLTS. DEVICE LONGEVITY WAS ESTIMATED AT 34 MONTHS BASED ON THERAPY MEASUREMENT OF 1000 OHMS. THE DEVICE HAD BEEN IMPLANTED FOR 36 MONTHS". ADDITIONAL INFO HAS BEEN REQUESTED. A FOLLOW-UP REPORT WILL BE SUBMITTED IF ADDITIONAL INFO BECOMES AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNERGY | LGW | MEDTRONIC PUERTO RICO OPERATIONS CO | 7427 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LEAD MODEL 3998| EXPLANTED| EXTENSION MODEL 7489| EXPLANTED| EXPLANTED| EXTENSION MODEL 7489 |