FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 11108397 · Received January 4, 2021

Report

Report Number
1000113657-2021-00002
Event Type
Malfunction
Date Received
January 4, 2021
Date of Event
December 10, 2020
Report Date
February 9, 2021
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

SECTIONS WITH ADDITIONAL INFORMATION AS OF 09-FEB-2021: D10/H3/H6: EXPECTED METER WAS RETURNED FOR EVALUATION. PRODUCT EVALUATION RESULTS OF REPORTED DEFECT NOT REPRODUCED ON RETURNED METER. CUSTOMER RETURNED INCORRECT STRIPS MOST LIKELY UNDERLYING ROOT CAUSE: MLC-058: USER HAD AN INACCURATE REFERENCE: SELF: THE PERSON IS USING THEMSELVES AS THE REFERENCE OR HOW THEY FEEL AT THE TIME THEY RUN THE BLOOD TEST. POOR TECH-INACCURATE REFERENCE.

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). PRODUCT IS NOT RETURNED FOR INVESTIGATION YET. NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 24-DEC-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER WHO STATED THE PRODUCT RESOLVED THE INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 242, 202, 199 AND 220 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 90-120 MG/DL AND EXPECTED POST-MEAL (3-4 HOURS) FASTING BLOOD GLUCOSE TEST RESULT IS 160 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BLOOD TEST WAS PERFORMED BY THE CUSTOMER FASTING (2 HOURS POST MEAL) AND PRODUCED TEST RESULT OF 232 MG/DL USING THE METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 02/13/2022 AND TEST STRIPS WERE OPENED THREE DAYS AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
5637 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX NFRS24/CS 50CT MG/DL ZX4016S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1