FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 11108355 · Received January 4, 2021

Report

Report Number
1911916-2020-01164
Event Type
Malfunction
Date Received
January 4, 2021
Date of Event
December 7, 2020
Report Date
January 8, 2021
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
00382903096534
PMA / PMN Number
K980987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY RECORD REVIEW WAS PERFORMED FOR PROVIDED LOT NUMBER 170298. THE REVIEW DID NOT REVEAL ANY DETECTED QUALITY ISSUES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS REPORTED INCIDENT. TO AID IN THE INVESTIGATION OF THIS INCIDENT, ONE SAMPLE WAS RECEIVED FOR EVALUATION BY OUR QUALITY TEAM. IT CAME IN AN OPENED PACKAGING BLISTER AND A VISUAL INSPECTION WAS PERFORMED. THE SYRINGE BARREL HAS EMBEDDED DEGRADED RESIN. THE CAUSE FOR THIS DEFECT COULD HAVE RESULTED DURING THE SYRINGE MOLDING PROCESS, WHERE THE EMBEDDED RESIN BUILDS UP IN THE HOT RUNNER SYSTEM AND CAN BECOME LOOSE, BREAK OFF AND BECOME MOLDED INTO THE COMPONENTS.

Description of Event or Problem · 0

IT WAS REPORTED THAT SYRINGE 50ML LL TIP 1ML HAD FOREIGN MATTER IN THE BARREL OF THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 309653, BATCH NO.: 0170298. IT WAS REPORTED THAT A BROWN STAIN OR FOREIGN OBJECT WAS FOUND ON AND IN THE BARREL OF THE SYRINGE. PER COMPLAINT FORM: A STAFF MEMBER OPENED THE PACKAGING OF A 50ML SYRINGE TO PREPARE FOR MIXING UP A MEDICATION. UPON OPENING THE PACKAGING IT WAS NOTED THAT THERE IS A BROWN STAIN ON AND IN THE BARREL OF THE SYRINGE. THE STAINING WAS NOT NOTICED UNTIL AFTER THE PACKAGING WAS OPENED.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 50ML LL TIP 1ML HAD FOREIGN MATTER IN THE BARREL OF THE SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO.: 309653 BATCH NO.: 0170298. IT WAS REPORTED THAT A BROWN STAIN OR FOREIGN OBJECT WAS FOUND ON AND IN THE BARREL OF THE SYRINGE. PER COMPLAINT FORM: A STAFF MEMBER OPENED THE PACKAGING OF A 50ML SYRINGE TO PREPARE FOR MIXING UP A MEDICATION. UPON OPENING THE PACKAGING IT WAS NOTED THAT THERE IS A BROWN STAIN ON AND IN THE BARREL OF THE SYRINGE. THE STAINING WAS NOT NOTICED UNTIL AFTER THE PACKAGING WAS OPENED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11268 SYRINGE 50ML LL TIP 1ML PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 0170298 00382903096534

Patients

Seq Age Sex Outcome Treatment
1