FDA Adverse Event Injury Summary report: N

GMK-SPHERE TIBIAL INSERT FIXED FLEX #4/10 MM R

MDR report key: 11106974 · Received January 4, 2021

Report

Report Number
3005180920-2021-00004
Event Type
Injury
Date Received
January 4, 2021
Date of Event
December 7, 2020
Report Date
January 4, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030826542
PMA / PMN Number
K121416
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CLINICAL EVALUATION PERFORMED BY MEDACTA MEDICAL AFFAIRS MANAGER: BILATERAL INSERT REVISION IN TKA 14 MONTHS AFTER PRIMARY FOR JOINT INSTABILITY. THE INSERT WAS EXCHANGED TO A THICKER ONE. THE DEVELOPMENTAL INSTABILITY IS A RATHER COMMONLY DESCRIBED POSSIBLE COMPLICATION FOLLOWING TKA, BECAUSE SOFT TISSUE MAY STRETCH AND PROVIDE INSUFFICIENT STABILITY. IT IS THEREFORE COMMON PRACTICE TO RESORT TO A THICKER INSERT AND RECREATE SOFT TISSUE TENSION. THIS IS NOT A PROBLEM DUE TO ANY IMPLANT DEFECT OR MALFUNCTION.

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 14 DECEMBER 2020: LOT 1902500: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-MAY-2019. EXPIRATION DATE: 2024-04-28. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT. ADDITIONAL DEVICE INVOLVED: BATCH REVIEW PERFORMED ON 14 DECEMBER 2020: LOT 186023: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 13-NOV-2018. EXPIRATION DATE: 2023-11-01. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

THE PATIENT CAME IN REPORTING INSTABILITY ON BOTH KNEES. THE SURGEON PERFORMED A BILATERAL POLY-SWAP AND REVISED THE RIGHT GMK-SPHERE TIBIAL INSERT - FLEX S4R - 10MM TO A GMK-SPHERE TIBIAL INSERT - FLEX S4R - 14MM AND THE LEFT SPHERE INSERT FLEX LEFT 11MM S4 TO A GMK-SPHERE TIBIAL INSERT - FLEX S4L - 14MM, TO GIVE THE PATIENT MORE STABILITY. THE SURGERY WAS COMPLETED SUCCESSFULLY, 1 YEAR AND 2 MONTHS AFTER PRIMARY SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9450 GMK-SPHERE TIBIAL INSERT FIXED FLEX #4/10 MM R KNEE TIBIAL INSERT JWH MEDACTA INTERNATIONAL SA 02.12.0410FR 1902500 07630030826542

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention