MDR report key: 1110688
·
Received August 14, 2008
Report
- Report Number
- 1824206-2008-00142
- Event Type
- Malfunction
- Date Received
- August 14, 2008
- Date of Event
- February 26, 2008
- Report Date
- February 26, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Removal / Correction Number
- Z-0240-2008
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
MAGNET HEATSTAKE ASSEMBLY WAS BROKEN. THE RIGHT FOOT MOUNT BRACKET WAS NOT CUT IN SPEC, SO THE SIDERAIL WOULD NOT LOCK IN UP POSITION. USED DEVICE AND REPLACED RIGHT FOOT MOUNT BRACKET. ALSO REPLACED MAGNET HEATSTAKE ASSEMBLY. NO OTHER PROBLEMS FOUND, NO INJURIES, BED BACK IN SERVICE.
Description of Event or Problem · 1
BED SIDERAIL WOULD NOT LOCK IN THE UP POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERSACARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | 3200 | NA |