FDA Adverse Event Malfunction Summary report: N

DIGNICAP DELTA

MDR report key: 11103344 · Received December 31, 2020

Report

Report Number
3012146457-2020-00002
Event Type
Malfunction
Date Received
December 31, 2020
Date of Event
July 16, 2020
Report Date
December 31, 2020
Manufacturer
DIGNITANA, INC
Product Code
PMC
PMA / PMN Number
K191166
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

HEAD SIZE OF THE PATIENTS USING SCALP COOLING THERAPY CAN VARY AMONG OUTLIER DEMOGRAPHICS, WHICH IS WHY WE RECOMMEND WEARING A HEADBAND ALONG WITH THE COOLING WRAP. THIS WILL HELP AVOID ANY CONTACT THE COLD SURFACE OF THE COOLING WRAP AND THE SKIN. THIS ADVERSE EVENT COULD HAVE BEEN AVOIDED IF THE PATIENT HAD WON THE HEADBAND DURING THE TREATMENT. AFTER BECOMING AWARE OF THE EVENT, WE HAVE INTRODUCED SEVERAL MEASURES TO COUNTER THIS ISSUE WHICH INCLUDES REQUIREMENT TO WEAR HEADBANDS DURING THE SCALP COOLING SESSION INTRODUCING SMALLER WRAP TO USER FACILITIES FOR OUTLIER PATIENTS USER TRAINING ON APPLICATION OF THE COOLING WRAP AND REDUCING COOLING INTENSITY OF THE DEVICE IF PATIENT HAS ALREADY GONE THROUGH HEAVY ALOPECIA AND HAS EXPOSED BALD SPOTS.

Description of Event or Problem · 1

PATIENT RECEIVED SKIN DARKENING ON THE FOREHEAD AREA AFTER THE SCALP COOLING SESSION WITH (B)(4) DEVICE. WE BELIEVE THIS OCCURED DUE TO DIRECT CONTACT ESTABLISHING BETWEEN THE COLD SURFACE OF THE COOLING WRAP AND BARE SKIN AROUND THESE AREAS. UPON FURTHER INVESTIGATION THROUGH COMMUNICATION/INTERVIEWS, IT WAS FOUND THAT PATIENT HAD NOT WORN HEADBAND IN THEIR 1ST TREATMENT. THIS CAUSED THEIR FOREHEAD SKIN TO BE EXPOSED AND IN CONTACT WITH THE COLD SURFACE OF THE COOLING WRAP FOR THE DURATION OF THE SCALP COOLING SESSION CAUSING THE CRYOGENIC BURNS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1565005 DIGNICAP DELTA DIGNICAP DELTA SCALP COOLING SYSTEM PMC DIGNITANA, INC

Patients

Seq Age Sex Outcome Treatment
1 40 YR Required Intervention