Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND REPLACED THE SRAM MEMORY. SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE SYSTEM IS DISPLAYING A CHECKSUM ERROR. NO PT INJURY WAS REPORTED.