FDA Adverse Event Malfunction Summary report: N

CADD

MDR report key: 11099191 · Received December 30, 2020

Report

Report Number
3012307300-2020-12890
Event Type
Malfunction
Date Received
December 30, 2020
Date of Event
December 1, 2020
Report Date
March 2, 2021
Manufacturer
ST PAUL
Product Code
FRN
UDI-DI
10610586022128
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

ONE CADD CASSETTE WAS RETURNED FOR ANALYSIS. LEAK TESTING ON THE SAMPLE RECEIVED WAS PERFORMED USING A SYRINGE TO LOOK FOR UNUSUAL FUNCTIONS. NO LEAK WAS DETECTED. THE COMPLAINT WAS NOT CONFIRMED. PRODUCTION PERSONNEL TURN TUBE BACK AND FORTH FOR A MINIMUM OF TWO (2) SECONDS. REMOVE IMMEDIATELY AFTER TWO SECONDS WITH A MAXIMUM OF FIVE SECONDS IN THE SOLVENT ACCORDING TO PROCEDURE. PRODUCTION JOIN ALL COMPONENTS IMMEDIATELY AFTER SOLVENT APPLICATION. PRODUCTION PERFORMS 100 PERCENT VISUAL INSPECTION TO ASSURE THAT THE PARTS SHALL NOT BE DAMAGED INCLUDING SCRATCHES) OR DISTORTED WARPED. PRODUCTION PERFORMS 100 PERCENT LEAK TEST TO ASSURE ASSEMBLIES HAVE BEEN MANUFACTURED IN ACCORDANCE TO PROCEDURE AND THE TUBE AND BAG COMPONENTS DOES NOT PRESENT A LEAK OVERALL. PER PROCEDURE QUALITY SAMPLE 15 UNITS AT AN INTERVAL OF TWO HOURS PLUS OR MINUS 30 MINUTES, PRIOR TO PLACING PRODUCT IN BAG IN ORDER TO VERIFY PARTS ARE FREE OF DAMAGE, SCUFFS, PINCH MARKS, ETC.), CRACKS, CRAZING, CUTS, OR OTHER WORKMANSHIP DEFECTS THAT CAN AFFECT ASSEMBLY APPEARANCE AND FUNCTION.

Additional Manufacturer Narrative · 1

MANUFACTURE DATE IS UNKNOWN SINCE THE SPECIFIC LOT IS UNKNOWN.

Description of Event or Problem · 1

INFORMATION WAS RECEIVED REGARDING A CADD CASSETTE RESERVOIR. IT WAS REPORTED THAT THE PRODUCT LEAKED INTO THE HOUSING CASSETTE. THIS OCCURRED DURING FILLING, SO IT HAD NOT LEFT THE PHARMACY DEPARTMENT. CUSTOMER IS NOT SURE OF THE AFFECTED LOT, BUT THEY REPORT THAT IT IS ONE OF TWO LOT NUMBERS (3840915 OR 18X068).