FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 11093730 · Received December 30, 2020

Report

Report Number
2016493-2020-80899
Event Type
Malfunction
Date Received
December 30, 2020
Report Date
September 27, 2019
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

(B)(4). CASE DESCRIPTION: (B)(6) HAD ERROR 600.6815 ON PCU SN (B)(4). (B)(6) WAS NOT ABLE TO TRANSFER NETWORK CONFIGURATION. ERROR MESSAGE: " AUTHENTICATION FAILED WHILE PERFORMING TRANSFER NETWORK CONFIGURATION TASK" FAILURE DEVICE TYPE: ALARIS SYSTEM INSTRUMENT. FAILURE PROBLEM TYPE: 8015. FAILURE MODE: TROUBLESHOOTING/ ERROR CODES. CASE RESOLUTION: I REMOTE IN TO (B)(6) COMPUTER AND FOUND OUT PCU MUST BE ZEROLIZED BECAUSE HE USED FIP ENABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1563292 8015 ALARIS SYSTEM PC UNIT PUMP, INFUSION FRN CAREFUSION SD 8015 10885403801518

Patients

Seq Age Sex Outcome Treatment
1