8015 ALARIS SYSTEM PC UNIT
Report
- Report Number
- 2016493-2020-80899
- Event Type
- Malfunction
- Date Received
- December 30, 2020
- Report Date
- September 27, 2019
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- UDI-DI
- 10885403801518
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER¿S REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A TRACKWISE COMPLAINT HISTORY REVIEW WAS COMPLETED, AND IT WAS CONFIRMED THAT THERE WERE ADDITIONAL COMPLAINTS RECEIVED WITH SIMILAR SN (B)(4) FOR THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). CASE DESCRIPTION: (B)(6) HAD ERROR 600.6815 ON PCU SN (B)(4). (B)(6) WAS NOT ABLE TO TRANSFER NETWORK CONFIGURATION. ERROR MESSAGE: " AUTHENTICATION FAILED WHILE PERFORMING TRANSFER NETWORK CONFIGURATION TASK" FAILURE DEVICE TYPE: ALARIS SYSTEM INSTRUMENT. FAILURE PROBLEM TYPE: 8015. FAILURE MODE: TROUBLESHOOTING/ ERROR CODES. CASE RESOLUTION: I REMOTE IN TO (B)(6) COMPUTER AND FOUND OUT PCU MUST BE ZEROLIZED BECAUSE HE USED FIP ENABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1563292 | 8015 ALARIS SYSTEM PC UNIT | PUMP, INFUSION | FRN | CAREFUSION SD | 8015 | 10885403801518 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |