FDA Adverse Event Malfunction Summary report: N

TANDEM X2 SLIM CONTROL IQ

MDR report key: 11093608 · Received December 29, 2020

Report

Report Number
MW5098561
Event Type
Malfunction
Date Received
December 29, 2020
Date of Event
December 15, 2020
Report Date
December 25, 2020
Manufacturer
TANDEM DIABETES CARE, INC.
Product Code
QJI
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

I AM TYPE 1 DIABETIC WEAR TANDEM X2 SLIM CONTROL IQ PUMP, ON MY FLIGHT FORM (B)(6) TO (B)(6) PUMP STOPPED INSULIN DELIVERIES AND FLASHED ALTITUDE ERROR. ON THEIR WEBSITE THERE IS NO WARNING OR WHAT TO DO IF YOU FLY. CALLED UP TANDEM AND CAME TO KNOW BECAUSE OF PRESSURE THIS HAPPENED AND THE 6 SMALL HOLES IN THE BACK GOT BLOCKED. THEY SHOULD HAVE THIS WARNING IN RED AND HOW TO PREVENT IT FROM HAPPENING. FDA SAFETY REPORT ID# (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1549912 TANDEM X2 SLIM CONTROL IQ INTEROPERABLE AUTOMATED GLYCEMIC CONTROLLER QJI TANDEM DIABETES CARE, INC. X2 SLIM

Patients

Seq Age Sex Outcome Treatment
1 59 YR Other