STEM: AMISTEM H HA COATED STD STEM SIZE 1
Report
- Report Number
- 3005180920-2020-00951
- Event Type
- Injury
- Date Received
- December 29, 2020
- Date of Event
- December 1, 2020
- Report Date
- February 18, 2021
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LZO
- UDI-DI
- 07630030804076
- PMA / PMN Number
- K093944
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- 003
Narratives
VISUAL INSPECTION PERFORMED BY MEDACTA R&D DEPARTMENT: VISUAL INSPECTION PERFORMED ON 19TH JANUARY 2021. NO PARTICULAR SIGNS WERE NOTICED ON THE STEM AND HEAD; SOME LITTLE SIGNS ON THE SPHERICAL SURFACE OF THE CERAMIC BALL HEAD ARE PROBABLY RELATED TO THE REVISION SURGERY, WHILE THE STEM WAS COMPLETELY UNCOVERED FROM HA, INDICATOR OF PROPER OSSEOINTEGRATION. FROM THE RECEIVED INFORMATION THE ROOT CAUSE OF THE EVENT REMAINS UNKNOWN, BUT ASEPTIC LOOSENING IS COMMON BETWEEN THE ADVERSE EVENTS DESCRIBED IN LITERATURE AFTER PRIMARY CEMENTLESS HIP ARTHROPLASTIES.
BATCH REVIEW PERFORMED ON 10 DECEMBER 2020 LOT 140745: 100 ITEMS MANUFACTURED AND RELEASED ON 23-APR-2014. EXPIRATION DATE: 2019-03-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE,ALL ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITH ONE SIMILAR EVENT REPORTED SINCE 2016.
THE SURGEON PERFORMED A REVISION (6 YEARS AND 3 MONTHS AFTER PRIMARY SURGERY) ON THE RIGHT HIP OF THE PATIENT DUE TO AN ASEPTIC LOOSENING. STEM AND HEAD HAVE BEEN SUCCESSFULLY REVISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1553321 | STEM: AMISTEM H HA COATED STD STEM SIZE 1 | CEMENTLESS HIP STEM | LZO | MEDACTA INTERNATIONAL SA | 01.18.131 | 140745 | 07630030804076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Required Intervention |