AS SYR PSD MICROBORE
Report
- Report Number
- 9616066-2020-20656
- Event Type
- Malfunction
- Date Received
- December 28, 2020
- Date of Event
- November 19, 2020
- Report Date
- January 15, 2021
- Manufacturer
- SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 50885403233904
- PMA / PMN Number
- K811885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H6: INVESTIGATION SUMMARY: NO PRODUCT OR PHOTO WAS RETURNED BY THE CUSTOMER. THE CUSTOMER COMPLAINT OF LEAK AT BASE OF CALVE AND TOP OF SYRINGE COULD NOT BE VERIFIED DUE TO THE PRODUCT NOT BEING RETURNED FOR FAILURE INVESTIGATION. THE ROOT CAUSE OF THIS FAILURE COULD NOT BE IDENTIFIED WITHOUT A FAILURE INVESTIGATION. A DEVICE HISTORY RECORD REVIEW FOR MODEL 10014914 LOT NUMBER 20015796 WAS PERFORMED. THE SEARCH SHOWED THAT A TOTAL OF (B)(4) UNITS IN 1 LOT NUMBER WAS BUILT ON 09JAN2020. THERE WERE NO QUALITY NOTIFICATIONS ISSUED FOR THE FAILURE MODE REPORTED BY THE CUSTOMER DURING THE PRODUCTION BUILD OF THIS SET. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT IV FLUID LEAKED FROM THE AS SYR PSD MICROBORE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IV FLUID FOUND ON FLOOR. LEAK OBSERVED AT BASE OF CLAVE AND TOP OF SYRINGE TUBING DURING INFUSION."
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT IV FLUID LEAKED FROM THE AS SYR PSD MICROBORE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "IV FLUID FOUND ON FLOOR. LEAK OBSERVED AT BASE OF CLAVE AND TOP OF SYRINGE TUBING DURING INFUSION."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1539306 | AS SYR PSD MICROBORE | INTRAVASCULAR ADMINISTRATION SET | FPA | SISTEMAS MEDICOS ALARIS, S.A. DE C.V. | 10014914 | 20015796 | 50885403233904 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |