FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II CLOSED IV CATHETER SYSTEM

MDR report key: 11084012 · Received December 28, 2020

Report

Report Number
3006948883-2020-01047
Event Type
Malfunction
Date Received
December 28, 2020
Date of Event
November 10, 2020
Report Date
December 30, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 0231222. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED AT THE NEEDLE TUBE JUNCTION DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: "ON (B)(6) 2020, AT 11:37 AM, THE CLOSED INTRAVENOUS INDWELLING NEEDLE WAS USED TO PERFORM INTRAVENOUS PUNCTURE SUCCESSFULLY, AND THE DISPOSABLE POUCH INFUSION DEVICE WAS CONNECTED WITH THE NEEDLE FOR TREATMENT. DURING THE FLUID INFUSION, THE NURSE ON DUTY FOUND THAT THE Y TYPE HOSE BASE OF THE CLOSED INTRAVENOUS INDWELLING NEEDLE WAS LEAKING AT THE JUNCTION WITH THE NEEDLE TUBE."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM LEAKED AT THE NEEDLE TUBE JUNCTION DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "ON (B)(6) 2020, AT 11:37 AM, THE CLOSED INTRAVENOUS INDWELLING NEEDLE WAS USED TO PERFORM INTRAVENOUS PUNCTURE SUCCESSFULLY, AND THE DISPOSABLE POUCH INFUSION DEVICE WAS CONNECTED WITH THE NEEDLE FOR TREATMENT. DURING THE FLUID INFUSION, THE NURSE ON DUTY FOUND THAT THE Y TYPE HOSE BASE OF THE CLOSED INTRAVENOUS INDWELLING NEEDLE WAS LEAKING AT THE JUNCTION WITH THE NEEDLE TUBE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1548831 BD INTIMA-II CLOSED IV CATHETER SYSTEM INTRAVASCULAR CATHETER FOZ BD (SUZHOU) 0231222

Patients

Seq Age Sex Outcome Treatment
1