SWAB ALCOHOL 100 BX 1200 ASIA PACIFIC
Report
- Report Number
- 1920898-2020-01761
- Event Type
- Malfunction
- Date Received
- December 28, 2020
- Date of Event
- November 24, 2020
- Report Date
- January 15, 2021
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- LKB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: CUSTOMER RETURNED (4) BD ALCOHOL SWABS (3 LOOSE, 1 SEALED) FROM LOT # 0163467. CUSTOMER STATES THAT THERE IS FOREIGN MATTER. THE RETURNED SWABS WERE EXAMINED AND 2 OUT OF 4 SWABS EXHIBITED SPLICING TAPE ON THE SWAB PAD. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH #0163467. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. ROOT CAUSE: ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT SWAB ALCOHOL 100 BX 1200 ASIA PACIFIC HAD FOREIGN MATTER ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOREIGN MATTER.
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SWAB ALCOHOL 100 BX 1200 ASIA PACIFIC HAD FOREIGN MATTER ON 2 OCCASIONS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: FOREIGN MATTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1548214 | SWAB ALCOHOL 100 BX 1200 ASIA PACIFIC | PAD, ALCOHOL, DEVICE DISINFECTANT | LKB | BD MEDICAL - DIABETES CARE | 0163467 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |