LACTATE DEHYDROGENASE (LDH)
Report
- Report Number
- 3002809144-2020-01212
- Event Type
- Malfunction
- Date Received
- December 28, 2020
- Date of Event
- December 9, 2020
- Report Date
- January 20, 2021
- Manufacturer
- ABBOTT GMBH
- Product Code
- CFJ
- UDI-DI
- 00380740161460
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A REVIEW OF SIMILAR COMPLAINT TICKETS COULD NOT BE PERFORMED AS THE LDH REAGENT LOT NUMBER WAS UNKNOWN. TRENDING REVIEW DETERMINED NO TRENDS FOR FALSELY ELEVATED RESULTS FOR THE PRODUCT. RETURN TESTING WAS NOT COMPLETED AS RETURNS WERE NOT AVAILABLE. THE PATIENT SAMPLE WAS RETESTED ON THE SAME INSTRUMENT AND ACCEPTABLE RESULTS WERE GENERATED. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES ASSOCIATED WITH THE CUSTOMER'S OBSERVATION. A REVIEW OF THE PRODUCT LABELING CONCLUDED THAT THE ISSUE IS SUFFICIENTLY ADDRESSED. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE LACTATE DEHYDROGENASE (LDH) REAGENT, LIST NUMBER 2P56 WAS IDENTIFIED.
AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THERE WAS NO ADDITIONAL PATIENT INFORMATION PROVIDED BY THE CUSTOMER DUE TO PRIVACY ISSUES.
THE CUSTOMER REPORTED FALSELY ELEVATED LDH RESULTS ON ONE PATIENT GENERATED ON THE ARCHITECT ANALYZER. THE RESULTS PROVIDED WERE: ON (B)(6) 2020 (B)(6) INITIAL LDH ON SERIAL NUMBER (B)(4) (NORMAL RANGE 125-220U/L) / RETEST ON SN (B)(4). THERE WAS NO REPORTED IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1542036 | LACTATE DEHYDROGENASE (LDH) | NAD REDUCTION/NADH OXIDATION, LACTATE DEHYDROGENASE | CFJ | ABBOTT GMBH | 2P56-22 | 00380740161460 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARC C16K PRC MOD, 03L77-01, C1600600| ARC C16K PRC MOD, 03L77-01, C1600600| ARC C16K PRC MOD, 03L77-01, C1600600 |