8100 ALARIS PUMP MODULE
Report
- Report Number
- 2016493-2020-81336
- Event Type
- Malfunction
- Date Received
- December 27, 2020
- Report Date
- November 5, 2019
- Manufacturer
- CAREFUSION SD
- Product Code
- FRN
- PMA / PMN Number
- K133532
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR (B)(6) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE (B)(6) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). DURING THE REPAIR PROCESS, IT WAS DETERMINED THAT THE ORIGINAL PROBLEM CODE SHOULD HAVE BEEN BROK (BROKEN DAMAGED) FOR COMPLAINT TRENDING PURPOSES FILE REOPENED TO ADD TRACK WISE PR NUMBER TO THE DEVICE DATA FIELD. THIS FILE WAS INITIALLY CLASSIFIED AS A LEVEL 3 NON-COMPLAINT FOR PREVENTATIVE MAINTENANCE, UPGRADE, RECONDITIONING, CUSTOMER INQUIRY, OR RECALL WHICH DO NOT REQUIRE CUSTOMER ADVOCACY QUALITY REVIEW OR A NO PATIENT INVOLVED STATEMENT. THIS FILE CONTAINED DOCUMENTATION OF PRODUCT DEFICIENCY ALLEGATIONS, AND/OR REPAIRS (WHICH MAY OR MAY NOT HAVE BEEN PERFORMED) THAT ARE OUT-OF-SCOPE WITH THE INITIAL LEVEL 3 NON-COMPLAINT. SUCH DOCUMENTATION UPGRADED THE FILE TO A LEVEL 2 COMPLAINT, WHICH REQUIRED A LEVEL 2 CUSTOMER ADVOCACY QUALITY REVIEW.
THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
LVP BEZEL POST RECALL GROUP 1 -DOM 2019- 08/26/2019 10:35:20 RICKY COLEMAN (RCOLEMAN) MARIBEL REGALADO BIOMED 760-940-7711 [email protected] 09/09/2019 13:19:47 NGOC T BUI (NBUI) EST - RCL TO MJR 09/12/2019 10:24:39 CRISLEIVY PENA (CRPENA) THERE WAS NO PATIENT INVOLVEMENT. CONFIRMED UPDATED FROM RCL TO MJR FOR THE MAJOR REPAIR NEEDED PER NGOC T BUI, SERVICE TECH. REPAIR APPROVED BY MARIBEL REGALADO, BIOMED, AT [email protected] FOR $365. NEW PO# 1093004. 09/28/2019 14:01:57 NGOC T BUI (NBUI) LVP BEZEL POST RECALL COMPLETED. REPLACED BROKEN BOTTOM REAR CASE, AND DAMAGED DOOR AT KEYPAD. REPLACED BROKEN AIL SENSOR RIGHT SIDE, AND COVER DOOR. REPLACED MOTOR HARNESS FLUID INGRESS. REPLACED CORRODED RIGHT,LEFT IUI, AND FLUID INGRESS FRAME MEMB. 10/08/2019 09:33:16 CHRISTINA CASTANEDA (CHCASTAN) 9632001960920413730700457111910418 11/05/2019 09:50:08 JOSEPH KUHLS (JKUHLS) DURING THE REPAIR PROCESS, IT WAS DETERMINED THAT THE ORIGINAL PROBLEM CODE SHOULD HAVE BEEN BROK (BROKEN DAMAGED) FOR COMPLAINT TRENDING PURPOSES FILE REOPENED TO ADD TRACK WISE PR NUMBER TO THE DEVICE DATA FIELD. THIS FILE WAS INITIALLY CLASSIFIED AS A LEVEL 3 NON-COMPLAINT FOR PREVENTATIVE MAINTENANCE, UPGRADE, RECONDITIONING, CUSTOMER INQUIRY, OR RECALL WHICH DO NOT REQUIRE CUSTOMER ADVOCACY QUALITY REVIEW OR A NO PATIENT INVOLVED STATEMENT. THIS FILE CONTAINED DOCUMENTATION OF PRODUCT DEFICIENCY ALLEGATIONS, AND/OR REPAIRS (WHICH MAY OR MAY NOT HAVE BEEN PERFORMED) THAT ARE OUT-OF-SCOPE WITH THE INITIAL LEVEL 3 NON-COMPLAINT. SUCH DOCUMENTATION UPGRADED THE FILE TO A LEVEL 2 COMPLAINT, WHICH REQUIRED A LEVEL 2 CUSTOMER ADVOCACY QUALITY REVIEW.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1538893 | 8100 ALARIS PUMP MODULE | PUMP, INFUSION | FRN | CAREFUSION SD | 8100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |