FDA Adverse Event Malfunction Summary report: N

TRUE METRIX

MDR report key: 11069554 · Received December 23, 2020

Report

Report Number
1000113657-2020-00915
Event Type
Malfunction
Date Received
December 23, 2020
Date of Event
November 25, 2020
Report Date
December 23, 2020
Manufacturer
TRIVIDIA HEALTH INC
Product Code
NBW
UDI-DI
00021292007836
PMA / PMN Number
K140100
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REPORT REFERENCE NUMBER: (B)(4). METER WAS RETURNED FOR EVALUATION; NO DEFECT WAS DETECTED. RETURNED TEST STRIPS WERE EXPIRED. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-012: PRODUCT EXPIRED NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL ON 11-DEC-2020 TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.

Description of Event or Problem · 1

CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS FROM RESULTS OBTAINED OF 168, 170, 308, 233 AND 179 MG/DL. THE CUSTOMER¿S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT IS 150 MG/DL. CUSTOMER WAS NOT USING THE PROPER TESTING TECHNIQUE. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. AT THE TIME OF THE CALL, THE CUSTOMER STATED HE WAS AT HIS DOCTORS OFFICE FOR A CHECKUP DUE TO HAVING COVID-19 AND NOT RELATED TO HIS DIABETES. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER¿S EXPIRATION DATE IS 09/19/2020 (EXPIRED) AND TEST STRIPS WERE OPENED ONE WEEK PRIOR TO CALL. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1532059 TRUE METRIX SYSTEM, TEST BLOOD GLUCOSE, OVER THE COUNTER NBW TRIVIDIA HEALTH INC STRIP, TRUE METRIX 100CTMG/DL MW3223S 00021292007836

Patients

Seq Age Sex Outcome Treatment
1