FDA Adverse Event Malfunction Summary report: N

DIGNICAP DELTA

MDR report key: 11065297 · Received December 23, 2020

Report

Report Number
3012146457-2020-75238
Event Type
Malfunction
Date Received
December 23, 2020
Date of Event
September 25, 2020
Report Date
December 18, 2020
Manufacturer
DIGNITANA, INC
Product Code
PMC
PMA / PMN Number
K191166
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

HEAD SIZE OF THE PATIENTS USING SCALP COOLING THERAPY CAN VARY AMONG OUTLIER DEMOGRAPHICS, WHICH IS WHY WE RECOMMEND WEARING A HEADBAND ALONG WITH THE COOLING WRAP. THIS WILL HELP AVOID ANY CONTACT BETWEEN THE COLD SURFACE OF THE COOLING WRAP AND THE SKIN. THIS ADVERSE EVENT COULD HAVE BEEN AVOIDED IF THE PATIENT HAD WORN THE HEADBAND DURING THE TREATMENT. AFTER THE BECOMING AWARE OF THE EVENT, WE HAVE INTRODUCED SEVERAL MEASURES TO COUNTER THIS ISSUE WHICH INCLUDES. REQUIREMENT TO WEAR HEADBANDS DURING THE SCALP COOLING SESSION. INTRODUCING SMALLER WRAP TO USER FACILITIES FOR OUTLIER PATIENTS. USER TRAINING ON APPLICATION OF THE COOLING WRAP AND REDUCING COOLING INTENSITY OF THE DEVICE. IF PATIENT HAS ALREADY GONE THROUGH HEAVY ALOPECIA AND HAS EXPOSED BALD SPOTS.

Description of Event or Problem · 1

PATIENT RECEIVED SKIN DARKENING ON THE FOREHEAD AND TEMPLE AREA AFTER THE SCALP COOLING SESSION WITH DIGNICAP DELTA DEVICE. WE BELIEVE THIS OCCURRED DUE TO DIRECT CONTACT ESTABLISHING BETWEEN THE COLD SURFACE OF THE COOLING WRAP AND BARE SKIN AROUND THESE AREAS. UPON FURTHER INVESTIGATION THROUGH COMMUNICATION/INTERVIEWS, IT WAS FOUND THAT PATIENT HAD WORN HEADBAND AT 1ST TREATMENT, BUT THE WAY THE HEADBAND WAS APPLIED COVERED HAIR AT HER TEMPLES. SO AT SECOND TREATMENT SHE CHOSE TO NOT WEAR HEADBAND. THIS CAUSED HER FOREHEAD AND TEMPLE SKIN TO BE EXPOSED AND IN CONTACT WITH THE COLD SURFACE OF THE COOLING WRAP FOR THE DURATION OF THE SCALP COOLING SESSION CAUSING THE CRYOGENIC BURNS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1533907 DIGNICAP DELTA DIGNICAP DELTA SCALP COOLING SYSTEM PMC DIGNITANA, INC

Patients

Seq Age Sex Outcome Treatment
1 33 YR Required Intervention