HEATER-COOLER SYSTEM 3T
Report
- Report Number
- 9611109-2020-00717
- Event Type
- Malfunction
- Date Received
- December 23, 2020
- Date of Event
- December 2, 2020
- Report Date
- February 8, 2021
- Manufacturer
- LIVANOVA DEUTSCHLAND
- Product Code
- DWC
- Removal / Correction Number
- Z-2076/2081-2015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H10: A REVIEW OF THE DHR COULD NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE OCCURRED ISSUE. THROUGH FOLLOW UP COMMUNICATION, LIVANOVA LEARNED THAT THE DEVICE WAS CLEANED REGULARLY AS PER THE IFU'S AND PLACED OUTSIDE OF THE OPERATION THEATER DURING USE. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
SEE INITIAL REPORT.
THERE WAS NO KNOWN PATIENT INVOLVEMENT. THE HEATER-COOLER 16-02-80 IS NOT DISTRIBUTED IN THE USA AND IT IS SIMILAR TO HEATER-COOLER 16-02-85, WHICH IS DISTRIBUTED IN THE USA (510(K) NUMBER: K191402). LIVANOVA (B)(4) IMPLEMENTED A FIELD SAFETY NOTICE FOR DISINFECTION AND CLEANING OF HEATER-COOLER DEVICES. THE Z NUMBER IS Z-2076/2081-2015. LIVANOVA (B)(4) MANUFACTURES THE HEATER-COOLER SYSTEM 3T . THE INCIDENT OCCURRED IN (B)(6). LIVANOVA INITIATED AN INVESTIGATION. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
LIVANOVA (B)(4) RECEIVED A REPORT THAT A HEATER-COOLER SYSTEM 3T DEVICE WAS FOUND TO BE CONTAMINATED WITH MYCOBACTERIUM CHIMAERA. THE LABORATORY REPORT CONFIRMING THE REPORTED CONTAMINATION HAS BEEN PROVIDED TO LIVANOVA. THERE IS NO KNOWN PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1531438 | HEATER-COOLER SYSTEM 3T | CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS | DWC | LIVANOVA DEUTSCHLAND | 16-02-80 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |