AFINION 2
Report
- Report Number
- 3003045237-2020-00010
- Event Type
- Malfunction
- Date Received
- December 23, 2020
- Date of Event
- December 15, 2020
- Report Date
- December 23, 2020
- Manufacturer
- ABBOTT DIAGNOSTICS TECHNOLOGIES AS
- Product Code
- JQT
- PMA / PMN Number
- K182988
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE EMPLOYEE REPORTED THAT WHEN THE INCIDENT HAPPENED THE POWER SUPPLY WAS CONNECTED TO THE WALL OUTLET BEFORE CONNECTING TO THE INSTRUMENT. THE USER MANUAL FOR THE INSTRUMENT INSTRUCTS THE USER TO CONNECT IN THE FOLLOWING ORDER: CONNECT THE POWER CABLE TO THE POWER CORD ADAPTER. INSERT THE PLUG FROM THE POWER CORD ADAPTER INTO THE POWER SOCKET IN THE BACK OF THE ANALYZER. PLUG THE POWER CORD TO WALL A OUTLET. THE ANALYZER IS IN TRANSIT TO THE MANUFACTURER AND FURTHER INVESTIGATION WILL BE PERFORMED WHEN THE ANALYZER HAS BEEN RETURNED.
AN ABBOTT EMPLOYEE REPORTED THAT THE ANALYZER DID NOT STAY POWERED ON. THE ANALYZER STARTED THE SELF TEST AND THEN SHUT OFF. THE EMPLOYEE DISCONNECTED AND RECONNECTED THE ANALYZER. WHEN CONNECTING THE POWER SUPPLY TO THE ANALYZER, SPARK WAS OBSERVED FROM THE BACK OF THE ANALYZER, IN THE PLUG-IN PORT, AND THERE WAS A BURNT SMELL. THE BLUE LIGHT IN THE POWER SUPPLY WAS ILLUMINATED. THE POWER SUPPLY WAS CONNECTED TO THE WALL OUTLET WHEN BEING RECONNECTED TO THE INSTRUMENT. THE EMPLOYEE REPORTED SHE WAS NOT HURT. THERE WAS NO INJURY AND NO EVIDENCE OF FIRE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1533882 | AFINION 2 | ANALYZER | JQT | ABBOTT DIAGNOSTICS TECHNOLOGIES AS | 1116554 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |