ARCOS CON SZ B STD 70MM
Report
- Report Number
- 0001825034-2020-04408
- Event Type
- Injury
- Date Received
- December 22, 2020
- Date of Event
- December 10, 2020
- Report Date
- February 11, 2021
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- LPH
- PMA / PMN Number
- K090757
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED.
UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS DETERMINED TO BE NOT REPORTABLE. THE REPORTED EVENT OF DEEP INFECTION >90 DAYS OCCURRED POST IMPLANTATION. DURING THE INVESTIGATION PROCESS A REVIEW OF THE STERILE CERTIFICATIONS WERE REVIEWED AND FOUND TO BE CONFORMING WITH NO APPLICABLE DEVIATIONS. DEVICES WERE VERIFIED TO HAVE GONE THROUGH ACCEPTABLE STERILIZATION PROCESS FOLLOWING ISO/AAMI/ASTM & EU PUBLISHED GUIDELINES. THERE ARE MULTIPLE FACTORS THAT MAY CONTRIBUTE TO AN INFECTION THAT ARE OUTSIDE THE CONTROL OF ZIMMER BIOMET, SUCH AS EXTERNAL FACTORS, I.E. HOSPITAL/SURGICAL ENVIRONMENT, PROVIDER RELATED RISK FACTORS, AND/OR PATIENT COMORBIDITIES/RISK FACTORS. AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUES IDENTIFIED AFFECTING IMPLANT SAFETY OR EFFECTIVENESS, THEREFORE IMPLANTED PRODUCTS ARE NOT IDENTIFIED AS THE SOURCE OR CONTRIBUTING TO THE REPORTED INFECTION. THE ROOT CAUSE OF THE REPORTED EVENT WAS DETERMINED TO BE UNRELATED TO THE IMPLANTED ZIMMER BIOMET DEVICE. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE DUE TO INFECTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
IT WAS REPORTED PATIENT UNDERWENT INITIAL HIP ARTHROPLASTY ON UNKNOWN DATE. SUBSEQUENTLY; THE PATIENT WAS REVISED DUE TO UNKNOWN REASON. THE PATIENT WAS LATER REVISED FOR A SECOND TIME APPROXIMATELY TEN MONTHS POST REVISION DUE TO INFECTION. ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.
(B)(4). CONCOMITANT MEDICAL PRODUCTS: UNKNOWN ARCOS DISTAL STEM, UNKNOWN FEMORAL HEAD. FOREIGN COUNTRY: (B)(4). CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, DUE TO THE LOCATION OF THE DEVICE IS UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE FOR UNKNOWN REASON. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1520949 | ARCOS CON SZ B STD 70MM | PROSTHESIS, HIP | LPH | ZIMMER BIOMET, INC. | N/A | 013250 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R | SEE H10 NARRATIVE. |