FDA Adverse Event Malfunction Summary report: N

T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY

MDR report key: 11042824 · Received December 19, 2020

Report

Report Number
3013756811-2020-147904
Event Type
Malfunction
Date Received
December 19, 2020
Date of Event
November 8, 2020
Report Date
December 19, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
OZO
UDI-DI
00853052007264
PMA / PMN Number
P180008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 200-420 MG/DL. REPORTEDLY, THE PUMP SUPPLIES WERE CHANGED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1507758 T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND OZO TANDEM DIABETES CARE 1000096 00853052007264

Patients

Seq Age Sex Outcome Treatment
1 35 YR