CRTS SM IL SG VLV ONLY
Report
- Report Number
- 3013886523-2020-00255
- Event Type
- Injury
- Date Received
- December 18, 2020
- Date of Event
- November 25, 2020
- Report Date
- December 1, 2020
- Manufacturer
- INTEGRA LIFESCIENCES SWITZERLAND SAR
- Product Code
- JXG
- PMA / PMN Number
- NI
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UNIQUE DEVICE IDENTIFIER (UDI): (B)(4). THE CERTAS VALVE WAS RETURNED FOR EVALUATION: DEVICE HISTORY RECORD (DHR) - LOT 4191136 CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK. FAILURE ANALYSIS - THE POSITION OF THE CAM WHEN VALVE WAS RECEIVED WAS AT SETTING 2. THE VALVE WAS VISUALLY INSPECTED; NEEDLE HOLES WERE NOTED IN THE NEEDLE CHAMBER. THE VALVE WAS HYDRATED. THE VALVE WAS LEAK TESTED AND ONLY LEAKED FROM THE NEEDLE HOLE IN THE NEEDLE CHAMBER. THE VALVE PASSED THE TEST FOR PROGRAMMING, OCCLUSION, REFLUX, SIPHON GUARD AND PRESSURE. NO ROOT CAUSE COULD BE DETERMINED AS THE TECHNICIAN COULD NOT CONFIRM ANY PROBLEM WITH THE VALVE AT THE TIME OF INVESTIGATION. THE POSSIBLE ROOT CAUSE FOR THE ISSUE REPORTED BY THE CUSTOMER COULD BE DUE BIOLOGICAL DEBRIS AND PROTEIN BUILDUP INTERFERING WITH THE VALVE MECHANISM AT THE TIME OF INVESTIGATION, NO OCCLUSION ISSUES WERE NOTED.
N/A.
AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
A PHYSICIAN REPORTED THAT ON (B)(6) 2020, THE CERTAS VALVE FLOW COULD NOT BE CONFIRMED AND THE PATIENT WAS VOMITING. THE VALVE WAS IMPLANTED TO THE PATIENT VIA ARACHNOID- PERITONEAL SHUNT ON (B)(6) 2020 WITH INITIAL SETTING 3. THE PATIENT'S SYMPTOMS GOT BETTER A BUT THERE WAS NO CHANGE ABOUT VENTRICULAR SIZE ON X-RAY. FIVE MONTHS AFTER VALVE PLACEMENT, THE PATIENT STARTED VOMITING. WHEN THE SHUNT CONTRAST WAS PERFORMED, THE FLOW COULD BE CONFIRMED ONLY BY HIGH PRESSURE. THEREFORE, THE VALVE WAS REPLACED WITH A NEW ONE(HAKIM VALVE) ON (B)(6) 2020.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1504065 | CRTS SM IL SG VLV ONLY | CERTAS PLUS SMALL & RA W/SG | JXG | INTEGRA LIFESCIENCES SWITZERLAND SAR | 4191136 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |