FDA Adverse Event
Malfunction
Summary report: N
T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY
MDR report key: 11040712
·
Received December 18, 2020
Report
- Report Number
- 3013756811-2020-148842
- Event Type
- Malfunction
- Date Received
- December 18, 2020
- Date of Event
- December 11, 2020
- Report Date
- December 18, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613762
- PMA / PMN Number
- K201214
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PER THE TANDEM USER GUIDE: CHECK YOUR PUMP¿S PERSONAL SETTINGS REGULARLY TO ENSURE THEY ARE CORRECT. INCORRECT SETTINGS CAN RESULT IN OVER DELIVERY OR UNDER DELIVERY OF INSULIN. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT TARGET BLOOD GLUCOSE (BG) SETTING WAS PROGRAMMED INCORRECTLY. DURING TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, SETTING WAS CORRECTED AND INSULIN DELIVERY SUCCESSFULLY RESUMED. CUSTOMER'S BG WAS 156 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1505407 | T:SLIM X2 INSULIN PUMP WITH CONTROL-IQ TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613762 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR |