INGEVITY MRI
Report
- Report Number
- 2124215-2020-25026
- Event Type
- Injury
- Date Received
- December 18, 2020
- Date of Event
- October 2, 2020
- Report Date
- June 24, 2021
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- NVN
- UDI-DI
- 00802526523335
- PMA / PMN Number
- P150012/S000
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
IT WAS REPORTED THAT BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) RECEIVED A CALL FROM THE FIELD REPRESENTATIVE ABOUT LOSS OF CAPTURE (LOC) ON THE TELEMETRY STRIPS HOURS AFTER THE DEVICE AND LEADS WERE IMPLANTED. TS REVIEWED THE ELECTROGRAM ON LATITUDE AND SAW NO ISSUES, BUT SAW ON THE TELEMETRY STRIPS THERE WAS POSSIBLE LOC OR FUSION BEATS WITH POTENTIALLY GREATER THAN TWO SECONDS OF ASYSTOLE. THE FIELD REPRESENTATIVE NOTED THAT THEY SAW NO ISSUES DURING THE POWER-ON SELF TEST (POST) AND THE DEVICE CHECKED OUT PERFECTLY. THE DEVICE AND RIGHT VENTRICULAR LEAD WERE LATER REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
IT WAS REPORTED THAT BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) RECEIVED A CALL FROM THE FIELD REPRESENTATIVE ABOUT LOSS OF CAPTURE (LOC) ON THE TELEMETRY STRIPS HOURS AFTER THE DEVICE AND LEADS WERE IMPLANTED. TS REVIEWED THE ELECTROGRAM ON LATITUDE AND SAW NO ISSUES, BUT SAW ON THE TELEMETRY STRIPS THERE WAS POSSIBLE LOC OR FUSION BEATS WITH POTENTIALLY GREATER THAN TWO SECONDS OF ASYSTOLE. THE FIELD REPRESENTATIVE NOTED THAT THEY SAW NO ISSUES DURING THE POWER-ON SELF TEST (POST) AND THE DEVICE CHECKED OUT PERFECTLY. THE DEVICE AND LEADS REMAIN IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1502960 | INGEVITY MRI | IMPLANTABLE LEAD | NVN | BOSTON SCIENTIFIC CORPORATION | 7732 | 1015242 | 00802526523335 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| R |