FDA Adverse Event Malfunction Summary report: N

8100 ALARIS PUMP MODULE

MDR report key: 11036380 · Received December 18, 2020

Report

Report Number
2016493-2020-73735
Event Type
Malfunction
Date Received
December 18, 2020
Report Date
March 8, 2018
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403810015
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER REPORTED PROBLEM WAS CONFIRMED. THE DEVICE WAS REPAIRED, RETESTED, AND RELEASED BACK TO THE CUSTOMER. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

LVP BEZEL POST RECALL 2017 BEZEL ASSY- DAMAGE - UNSPECIFIED, IUI- DAMAGE- OTHER AND NON SUPPORTED PART INSTALLED DAMAGED/CRACKED- AIL SENSOR ASSY AND CASE REAR LVP BEZEL POST RECALL 2017- 03/08/2018 11:51:46 SANDRA J MCDONALD (SMCDONAL) LVP BEZEL POST RECALL 2017 JAMES HARLING 864-560-7867 EMAIL: [email protected] 03/19/2018 14:51:12 THONG TRANG (TTRANG) ESTIMATE MJR WAITING FOR APPROVAL. 03/22/2018 07:41:20 JESSICA GALANG (JGALANG) UPDATED FROM RCL TO MJR FOR THE MAJOR REPAIR NEEDED PER THONG TRANG, SERVICE TECH. REPAIR APPROVED BY RAY HARLING AT [email protected] FOR $365. NEW PO# IS 1123101 BILL. THERE WAS NO PATIENT INVOLVEMENT 03/22/2018 11:24:14 THONG TRANG (TTRANG) MISSED TAT MJR 03/23/2018 07:53:24 ANNETTE A MENDEZ (AMENDEZ) 1Z9791540272135631 08/05/2019 15:37:38 JOSEPH KUHLS (JKUHLS) DURING THE REPAIR PROCESS, IT WAS DETERMINED THAT THE ORIGINAL PROBLEM CODE SHOULD HAVE BEEN BROK (BROKEN DAMAGED) FOR COMPLAINT TRENDING PURPOSES FILE REOPENED TO ADD TRACK WISE PR NUMBER TO THE DEVICE DATA FIELD. THIS FILE WAS INITIALLY CLASSIFIED AS A LEVEL 3 NON-COMPLAINT FOR PREVENTATIVE MAINTENANCE, UPGRADE, RECONDITIONING, CUSTOMER INQUIRY, OR RECALL WHICH DO NOT REQUIRE CUSTOMER ADVOCACY QUALITY REVIEW, PER SWI 1501-006-000. THIS FILE CONTAINED DOCUMENTATION OF PRODUCT DEFICIENCY ALLEGATIONS, PER SWI-151-070-000 AND SWI 1501-096-000, AND/OR REPAIRS (WHICH MAY OR MAY NOT HAVE BEEN PERFORMED) THAT ARE OUT-OF-SCOPE WITH THE INITIAL LEVEL 3 NON-COMPLAINT. SUCH DOCUMENTATION UPGRADED THE FILE TO A LEVEL 2 COMPLAINT, PER SWI 1501-070-000 AND SWI-096-000, WHICH REQUIRED A LEVEL 2 CUSTOMER ADVOCACY QUALITY REVIEW, ALSO PER SWI 1501-070-000.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1502295 8100 ALARIS PUMP MODULE PUMP, INFUSION FRN CAREFUSION SD 8100 10885403810015

Patients

Seq Age Sex Outcome Treatment
1