FDA Adverse Event Injury Summary report: N

IC-BIPOLAR HEAD COCRMO Ø28/46

MDR report key: 11035225 · Received December 18, 2020

Report

Report Number
3012523063-2020-00014
Event Type
Injury
Date Received
December 18, 2020
Date of Event
August 20, 2019
Report Date
December 18, 2020
Manufacturer
IMPLANTCAST GMBH
Product Code
LZO
PMA / PMN Number
K180263
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

SINCE THE IMPLANTCAST GMBH WAS PROVIDED WITH NEITHER THE PRODUCTS, NOR PICTURES AN OPTICAL INSPECTION WAS NOT POSSIBLE. FOR THE SAME REASONS AN ASSESSMENT OF THE DIMENSIONAL ACCURACY AS WELL AS A FUNCTIONAL TESTING AND AN EXTERNAL SURVEY COULD NOT BE CARRIED OUT. THE PRODUCTION DOCUMENTS AND MATERIAL CERTIFICATIONS ARE FLAWLESS AND SHOW NO DEVIATIONS. FURTHERMORE, THE IMPLANTCAST GMBH WAS PROVIDED WITH AN X-RAY THAT WAS MADE PRIOR TO THE REVISION. IT IS VISIBLE THAT THE HEAD IS NOT ONLY DISLOCATED BUT ALSO DISCONNECTED. IT IS SUSPECTED THAT THE DISLOCATION HAPPENED FIRST AND AFTERWARDS THE DISCONNECTION. IT IS FURTHER SUSPECTED THAT THE DISLOCATION IS PATIENT-RELATED BECAUSE TO OUR KNOWLEDGE THERE IS NO PRODUCT FAILURE OR TECHNICAL ERROR RECOGNISABLE, BUT THIS CANNOT BE ESTABLISHED WITH ABSOLUTE CERTAINTY.

Description of Event or Problem · 1

ON (B)(6) 2020 THE IMPLANTCAST GMBH RECEIVED THE MESSAGE THAT AN IC- BIPOLAR HEAD DISLOCATED OUT OF THE NATURAL ACETABULUM AFTER BEING IMPLANTED FOR ABOUT 1 YEAR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1504148 IC-BIPOLAR HEAD COCRMO Ø28/46 LZO IMPLANTCAST GMBH 21510046

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R IC-HEAD BIOLOX® REF:25872800 LOT: 7011227992