FDA Adverse Event Malfunction Summary report: N

8015 ALARIS SYSTEM PC UNIT

MDR report key: 11034062 · Received December 18, 2020

Report

Report Number
2016493-2020-72163
Event Type
Malfunction
Date Received
December 18, 2020
Report Date
July 16, 2019
Manufacturer
CAREFUSION SD
Product Code
FRN
UDI-DI
10885403801518
PMA / PMN Number
K133532
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE TSC TROUBLESHOOTING PROCESS. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS NOT RETURNED FOR SERVICING WHICH NOT CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.

Description of Event or Problem · 1

CASE #: (B)(4), CASE SUBJECT: NPI 8015 NOT CONNECTING TO SYSTEMS MAINTENANCE ACCOUNT NAME: WEED ARMY COMMUNITY HOSPITAL ACCOUNT #: 1021395 ASSET NAME: 8015LS PCU DOM V9.19.1.2 A/B/G/N ASSET LOCATION: CONTACT: (B)(6) CONTACT EMAIL: (B)(6) CONTACT PHONE: (B)(6) CONTACT MOBILE: PATIENT OR USER INVOLVEMENT: NO PATIENT OR USER HARM: NO CASE DESCRIPTION: CUSTOMER REPORTS THEIR 8015 (SN: (B)(4)) NOT CONNECTING TO SYSTEMS MAINTENANCE. FAILURE DEVICE TYPE: ALARIS SYSTEM INSTRUMENT FAILURE PROBLEM TYPE: 8015 FAILURE MODE: TROUBLESHOOTING/ ERROR CODES CASE RESOLUTION: RECOMMENDED CUSTOMER CONNECT ANOTHER 8015 TO VERIFY PROPER SET UP. NEXT I INFORMED THE CUSTOMER THIS IS MOST LIKELY DUE TO THE SERIAL INPUT OUTPUT BOARD. RECOMMENDED REMOVING A KNOWN GOOD BOARD AND SWAPPING IN THE AFFECTED 8015. IF UNIT CONNECTS, REPLACE THE BOARD. IF UNIT STILL DOES NOT CONNECT THEN THEY ARE LOOKING AT A LOGIC BOARD ISSUE. PROVIDED PART NUMBER OF SIO BOARD. CUSTOMER WILL MOVE FORWARD WITH MY RECOMMENDATIONS. ENDED CALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1499245 8015 ALARIS SYSTEM PC UNIT PUMP,INFUSION FRN CAREFUSION SD 8015 10885403801518