FDA Adverse Event Injury Summary report: N

PUMP MMT-522NAS PRDGM INS V2.2 SK EN

MDR report key: 1103209 · Received August 8, 2008

Report

Report Number
3004209178-2008-00594
Event Type
Injury
Date Received
August 8, 2008
Date of Event
July 24, 2008
Report Date
July 27, 2008
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
LZG
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED DUE TO HYPERGLYCEMIA. THE CUSTOMER STATED THAT WHEN SHE REMOVED THE INFUSION SET, SHE DISCOVERED A FAT DEPOSIT IN THE CANNULA. TROUBLESHOOTING WAS PERFORMED, AND THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. DURING THE PRIME TEST, THE CUSTOMER SAW INSULIN EXIT THE TUBING AND THE INSULIN PUMP ALARMED NO DELIVERY. THE INSULIN PUMP PASSED THE HIGH PRESSURE TEST. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAS PRDGM INS V2.2 SK EN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAS

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization