FDA Adverse Event
Injury
Summary report: N
PUMP MMT-522NAS PRDGM INS V2.2 SK EN
MDR report key: 1103209
·
Received August 8, 2008
Report
- Report Number
- 3004209178-2008-00594
- Event Type
- Injury
- Date Received
- August 8, 2008
- Date of Event
- July 24, 2008
- Report Date
- July 27, 2008
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- LZG
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
THE CUSTOMER STATED THAT SHE WAS HOSPITALIZED DUE TO HYPERGLYCEMIA. THE CUSTOMER STATED THAT WHEN SHE REMOVED THE INFUSION SET, SHE DISCOVERED A FAT DEPOSIT IN THE CANNULA. TROUBLESHOOTING WAS PERFORMED, AND THE PROGRAMMING ON THE INSULIN PUMP WAS CORRECT. DURING THE PRIME TEST, THE CUSTOMER SAW INSULIN EXIT THE TUBING AND THE INSULIN PUMP ALARMED NO DELIVERY. THE INSULIN PUMP PASSED THE HIGH PRESSURE TEST. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAS PRDGM INS V2.2 SK EN | INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |