FDA Adverse Event Malfunction Summary report: N

SOCKET WRENCH FOR VEPTR NUT

MDR report key: 11028964 · Received December 17, 2020

Report

Report Number
2939274-2020-05634
Event Type
Malfunction
Date Received
December 17, 2020
Report Date
November 23, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HXC
UDI-DI
10705034720128
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

REPORTER IS A J&J EMPLOYEE. INVESTIGATION SUMMARY: SERVICE AND REPAIR EVALUATION: THE CUSTOMER REPORTED THAT THE DEVICE FAILED CALIBRATION. THE REPAIR TECHNICIAN REPORTED THE DEVICE FAILED TORQUE TESTING. THE CAUSE OF THE ISSUE IS FAILED HIGH. THE ITEM WILL BE REPAIRED AND WILL BE RETURNED TO THE CUSTOMER UPON COMPLETION OF THE SERVICE AND REPAIR PROCESS. ATTACHED SERVICE RECORD ROUTER COMPLETED THROUGH OPERATION 20. FINALIZED SERVICE RECORD WILL BE ARCHIVED IN TUNGSTEN DOCUMENT MANAGEMENT SYSTEM. THE EVALUATION WAS CONFIRMED. THE DEVICE WAS DEEMED SERVICEABLE AND WILL BE RETURNED TO THE CUSTOMER, NO DESIGN OR MANUFACTURING ISSUES WERE IDENTIFIED THEREFORE IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. DEVICE HISTORY LOT: PART NUMBER:03.641.004, SYNTHES LOT NUMBER: H852117-14, SUPPLIER LOT NUMBER: N/A, RELEASE TO WAREHOUSE DATE: JAN 23, 2020, EXPIRATION DATE: N/A, SUPPLIER: (B)(4). NO NCRS WERE GENERATED DURING PRODUCTION. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED ON (B)(6) 2020, DURING EVALUATION AT SERVICE AND REPAIR, A SOCKET WRENCH FOR VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB NUT FAILED IN CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SOCKET WRENCH FOR VEPTR NUT. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1491700 SOCKET WRENCH FOR VEPTR NUT WRENCH HXC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 03.641.004 H852117-14 10705034720128

Patients

Seq Age Sex Outcome Treatment
1