FDA Adverse Event
Malfunction
Summary report: N
CARESCAPE CENTRAL STATION
MDR report key: 11027135
·
Received December 17, 2020
Report
- Report Number
- 11027135
- Event Type
- Malfunction
- Date Received
- December 17, 2020
- Date of Event
- November 16, 2020
- Report Date
- December 9, 2020
- Manufacturer
- GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES, INC.
- Product Code
- DQK
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- CT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
STAFF REPORTED THAT THE CENTRAL STATION MADE A LOUD POPPING NOISE AND THEN THE SCREEN WENT BLACK. THIS RESULTED IN A LOSS OF CENTRALIZED PATIENT MONITORING. THE CENTRAL MONITOR WAS REPLACED AND UPON FURTHER INSPECTION, AN ISSUE WITH THE POWER SUPPLY WAS IDENTIFIED. THE MANUFACTURER RESPONDED TO THIS EVENT. GE FIELD ENGINEER HAS BEEN ON-SITE TO EVALUATE THE POWER SUPPLY AND BEGIN INVESTIGATION ON ROOT CAUSE. GE HAS PROVIDED INTERIM LOANER CENTRAL STATION UNTIL THE ISSUE IS RESOLVED AND HOSPITAL OWNED CENTRAL STATIONS ARE FIXED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1493652 | CARESCAPE CENTRAL STATION | COMPUTER, DIAGNOSTIC, PROGRAMMABLE | DQK | GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES, INC. | V2 INTEGRATED MAI 700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |