8300 ALARIS ETCO2 MODULE
Report
- Report Number
- 2016493-2020-70415
- Event Type
- Malfunction
- Date Received
- December 16, 2020
- Report Date
- June 13, 2019
- Manufacturer
- CAREFUSION SD
- Product Code
- CCK
- UDI-DI
- 10885403830013
- PMA / PMN Number
- K031741
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
THIS REPORTED EVENT AND SUBSEQUENT REPAIRS WERE INVESTIGATED THROUGH THE SERVICE REPAIR PROCESS. FAILURE DATA AND PARTS-USED INFORMATION WERE REVIEWED FOR THE SAP AND TRACK WISE FILES AND FOUND RELEVANT TO THE SERVICE REPAIR. A REVIEW OF THE DEVICE SERVICE HISTORY RECORD WAS PERFORMED FROM THE DATE OF MANUFACTURE TO THE DATE CORRESPONDING TO THIS SERVICE NOTIFICATION NUMBER. THE DATABASE SHOWED NO QUALITY NOTIFICATIONS WERE OPENED FOR THE DEVICE. A REVIEW OF THE DEVICE HISTORY RECORD IN SAP FOR SN (B)(4) WAS PERFORMED FROM THE DATE OF THE MANUFACTURE TO DATE OF THE RELEASE OF PRODUCT, WHICH CONFIRMED THAT THIS DEVICE WAS NOT INVOLVED IN A PRODUCTION FAILURE, AND PRODUCT WAS RETURNED FOR SERVICING WHICH CORRELATES TO THE CUSTOMER REPORTED ISSUE. A REVIEW OF THE COMPLAINT HISTORY RECORD IN THE TRACKWISE WAS PERFORMED FOR THE SN (B)(4) WHICH CONFIRMED NO SIMILAR COMPLAINTS WITH THE SAME OR RELATED FAILURE MODE. THE CUSTOMER STATED THAT THERE WAS NO PATIENT INVOLVEMENT.
(B)(4). UPDATED FROM MNR TO MJR FOR THE MAJOR REPAIR NEEDED PER LUAN LUONG, SERVICE TECH. REPAIR APPROVED BY CHRIS MCDERMOTT, CLINICAL ENGINEER AT [email protected] FOR $1625 FOR A NEW ORIDION BOARD. NO CHANGES TO PO#. (B)(4). ERROR 511 .2010 CAUSE BY U7 SOCKET. USE CO # 513743 TO FIX PROBLEM (U7 SOCKET). WE DO NOT HAVE TOOL TO REMOVE SOCKET . (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1484107 | 8300 ALARIS ETCO2 MODULE | PUMP, INFUSION | CCK | CAREFUSION SD | 8300 | 10885403830013 |