FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 11022786 · Received December 16, 2020

Report

Report Number
8010047-2020-10493
Event Type
Malfunction
Date Received
December 16, 2020
Date of Event
March 16, 2020
Report Date
December 16, 2020
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
UDI-DI
04953170363498
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE OLYMPUS SALES REP WAS ONSITE AND CONFIRMED THE CUSTOMER'S COMPLAINT. THE SALES REP TEMPORARILY USED THE COMP OUT TO THE OEV-262H MONITOR AND THEN USED SDI TO THE SECONDARY MONITOR. BASED ON THE INVESTIGATION, THE EVENT POINTED OUT WAS ESTIMATED TO BE CAUSED BY THE MALFUNCTION OF THE BOARD FOR PROCESSING THE IMAGE SIGNALS. THE MONITOR WAS NOT RETURNED TO OLYMPUS FOR REPAIR.

Description of Event or Problem · 1

AN OLYMPUS SALES REP REPORTED THAT HE TRIED TO CONNECT THE SDI CABLE FROM THE OTV-S190 PROCESSOR TO THE SDI 1 AND THE SDI 2 PORTS ON THE OEV-262H MONITOR BUT WAS UNABLE TO GET AN IMAGE. THE SALES REP WAS ABLE TO GET AN IMAGE TO DISPLAY ON THE MONITOR WHEN USING THE COMP OUT TO THE VIDEO IN ON THE OEV-262H MONITOR. THIS DEVICE MALFUNCTION OCCURRED DURING PREPARATION FOR USE AND THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1489718 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET OLYMPUS MEDICAL SYSTEMS CORP. OEV262H 04953170363498

Patients

Seq Age Sex Outcome Treatment
1