FDA Adverse Event
Malfunction
Summary report: N
BIPOL LEAD MODEL 300
MDR report key: 1102004
·
Received August 6, 2008
Report
- Report Number
- 1644487-2008-01788
- Event Type
- Malfunction
- Date Received
- August 6, 2008
- Date of Event
- April 28, 2008
- Report Date
- July 7, 2008
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY.
Description of Event or Problem · 1
REVIEW OF DIAGNOSTIC HISTORY FROM THE SITES HANDHELD COMPUTER REVEALED THAT THIS PATIENTS' DEVICE HAD HIGH LEAD IMPEDANCE FROM A SYSTEM DIAGNOSTIC TEST, END OF SERVICE STATUS WAS SET TO NO. REVIEW OF PROGRAMMING HISTORY INDICATES THAT THE DEVICE REMAINED ON WHEN THE HIGH LEAD IMPEDANCE WAS OBSERVED. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION FROM THE SITE ARE UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIPOL LEAD MODEL 300 | LYJ | CYBERONICS, INC. | 300-20 | 6850C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |