FDA Adverse Event Injury Summary report: N

TALENT THORACIC STENT GRAFT SYSTEM

MDR report key: 1101772 · Received August 7, 2008

Report

Report Number
2953200-2008-00658
Event Type
Injury
Date Received
August 7, 2008
Report Date
July 16, 2008
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070007
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION: RESULTS: (ARTERIAL TRAUMA/DISSECTION), (ACUTE TYPE B DISSECTION). CONCLUSION: (ACUTE TYPE B DISSECTION).

Description of Event or Problem · 1

A TALENT THORACIC STENT GRAFT SYSTEM WAS IMPLANTED FOR THE ENDOVASCULAR TREATMENT OF AN ACUTE TYPE B DISSECTION. THIS EVENT WAS REPORTED IN THE JOURNAL OF VASCULAR SURGERY 2008; 47: 1195-1202, 2008. THE STENT GRAFTS WERE IMPLANTED BETWEEN 2001 AND 2007. IT WAS REPORTED THE PT REC'D A TALENT DEVICE TO TREAT AN ACUTE TYPE B DISSECTION OCCURRED ON THE 3RD DAY AFTER IMPLANTATION CAUSED BY BARE SPRING-INDUCED INTIMAL INJURY. THORACIC ENDOVASCULAR REPAIR (TEVAR) INDICATION WAS PROGRESSIVE DILATATION. THE PT WAS TREATED BY AN EMERGENCY PROCEDURE INVOLVING SUPRA-CORONARY ARCH REPLACEMENT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PT WAS DISCHARGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT THORACIC STENT GRAFT SYSTEM MIH MEDTRONIC CARDIOVASCULAR NA UNK

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention