ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM
Report
- Report Number
- 2028159-2008-00283
- Event Type
- Injury
- Date Received
- August 7, 2008
- Date of Event
- July 8, 2008
- Report Date
- July 8, 2008
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- MPA
- PMA / PMN Number
- K951627
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER HAS QUARANTINED THE SYSTEM. THE SURGEON STATED HIS PARTNER NOTICED THAT THE SYSTEM BULB EXPIRED WHILE PERFORMING A CASE AN THE CASE WAS THEN COMPLETED USING THE ACCURUS HALOGEN SOURCE. SHORTLY THEREAFTER THE BULB WAS CHANGED BY THE IN HOUSE SERVICING BIOMEDICAL DEPARTMENT (GE), WITHOUT HIS AWARENESS. THE SURGEON IS REQUESTING THE ALCON PERFORM SERVICING AND INVESTIGATION OF THE UNIT PRIOR TO RISK MANAGEMENT EVAL SINCE THEY MAY POTENTIALLY CAUSE THE INVESTIGATION TO BE INACCURATE AS A RESULT. INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY WHEN ADD'L REPORTABLE INFO BECOMES AVAILABLE. THIS REPORT WAS MAILED TO FDA ON: 8/7/2008.
THE SURGEON REPORTED THAT ON A POST OPERATIVE EXAMINATION THE PT EXHIBITED SIGNS OF POTENTIAL PHOTOTOXICITY. THE SURGEON DESCRIBED A WELL CIRCUMSCRIBED 4-5 DISC DIAMETER TEMPORAL PARAFOVEAL PIGMENTARY DISRUPTION TO THE RETINAL PIGMENT EPITHELIUM. THE PT ALSO EXHIBITED DECREASE IN BEST CORRECTED VISUAL ACUITY. THE SURGEON HAS BEEN OPERATING UNDER 80% LIGHT OUTPUT THROUGHOUT ALL OF HIS CASES. HE IS UNAWARE OF THE LUMEN OUTPUT THAT IS EMITTED FROM THE LIGHT SOURCE. HE RECENTLY WAS INFORMED THAT BULB CHANGE WAS PERFORMED BY THEIR IN HOUSE SERVICING BIOMEDICAL DEPARTMENT (GE) AROUND THE TIME WHICH HE REPORTS THE PHOTOTOXICITY EVENT. THE SURGEON STATED THE PT'S POTENTIAL OUTCOME IS GUARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM | ENDOILLUMINATOR | MPA | ALCON - IRVINE TECHNOLOGY CENTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | Disability |