STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 8
MDR report key: 11009860
·
Received December 15, 2020
Report
- Report Number
- 3005180920-2020-00894
- Event Type
- Injury
- Date Received
- December 15, 2020
- Date of Event
- November 16, 2020
- Report Date
- December 15, 2020
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- JDI
- UDI-DI
- 07630030802096
- PMA / PMN Number
- K082792
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
BATCH REVIEW PERFORMED ON 27 NOVEMBER 2020: LOT 2002652: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-JUL-2020. EXPIRATION DATE: 2025-07-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.
Description of Event or Problem · 1
4 DAYS AFTER THE PRIMARY SURGERY, THE STEM SLID A LITTLE BIT DEEPER INTO THE BONE. ON X-RAYS A LONGITUDINAL FRACTURE OF THE FEMUR WAS DISCOVERED. THE SURGEON EXPLANTED THE STEM AND HE DID 5 CERCLAGES AROUND THE FEMUR AND IMPLANTED A QUADRA R STEM SIZE 6.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1478357 | STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 8 | CEMENTLESS HIP STEM | JDI | MEDACTA INTERNATIONAL SA | 01.12.028 | 2002652 | 07630030802096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |