FDA Adverse Event Injury Summary report: N

STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 8

MDR report key: 11009860 · Received December 15, 2020

Report

Report Number
3005180920-2020-00894
Event Type
Injury
Date Received
December 15, 2020
Date of Event
November 16, 2020
Report Date
December 15, 2020
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JDI
UDI-DI
07630030802096
PMA / PMN Number
K082792
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 27 NOVEMBER 2020: LOT 2002652: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 24-JUL-2020. EXPIRATION DATE: 2025-07-22. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

4 DAYS AFTER THE PRIMARY SURGERY, THE STEM SLID A LITTLE BIT DEEPER INTO THE BONE. ON X-RAYS A LONGITUDINAL FRACTURE OF THE FEMUR WAS DISCOVERED. THE SURGEON EXPLANTED THE STEM AND HE DID 5 CERCLAGES AROUND THE FEMUR AND IMPLANTED A QUADRA R STEM SIZE 6.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1478357 STEM: QUADRA-H CEMENTLESS, HA COATED STD STEM SIZE 8 CEMENTLESS HIP STEM JDI MEDACTA INTERNATIONAL SA 01.12.028 2002652 07630030802096

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention